502/Repairs: Difference between revisions

Jump to navigation Jump to search
No edit summary
No edit summary
Line 15: Line 15:
:'''Solution:''' Replace C180B-D
:'''Solution:''' Replace C180B-D


==Burned-out HV transformer==
:'''Symptom:''' HV transformer showing burns
:'''Root cause:''' overload by earlier V800 failure caused by leaky HV caps
:'''Solution:''' Replaced by 120-114 transformer (502) with 120-150 (from 502A) using only one cathode secondary.  Note the grid winding has more turns on the 120-150, so a dropper resistor needs to be added to prevent the grid going too negative (no beam).


[[Category:Instrument repair reports]]
[[Category:Instrument repair reports]]