:'''Solution:''' Replace C180B-D | :'''Solution:''' Replace C180B-D | ||
==Burned-out HV transformer== | |||
:'''Symptom:''' HV transformer showing burns | |||
:'''Root cause:''' overload by earlier V800 failure caused by leaky HV caps | |||
:'''Solution:''' Replaced by 120-114 transformer (502) with 120-150 (from 502A) using only one cathode secondary. Note the grid winding has more turns on the 120-150, so a dropper resistor needs to be added to prevent the grid going too negative (no beam). | |||