The second step is to get the timing (horizontal) unit working.  The best
The second step is to get the timing (horizontal) unit working.  The best
technique here is to apply a pulse train (100mV, 1MHz, for example) waveform  
technique here is to apply a pulse train (100 mV, 1 MHz, for example) waveform  
to the external trigger input of the timing unit while  
to the external trigger input of the timing unit while  
observing its outputs on a second oscilloscope.
observing its outputs on a second oscilloscope.
Some timing plug-ins such as the 3T77 provide a trigger pulse output on the
 
front panel.  Other plug-ins such as the 5T1A and 5T3 do not provide this
Some timing plug-ins such as the [[3T77]] provide a trigger pulse output on the
front panel.  Other plug-ins such as the [[5T1A]] and [[5T3]] do not provide this
signal on the front panel, so for those you have to probe the signal inside
signal on the front panel, so for those you have to probe the signal inside
the plug-in.  The timing plug-in is supposed to produce strobe pulses that
the plug-in.  The timing plug-in is supposed to produce strobe pulses that
Line 40: Line 41:


The third step is to get the vertical sampling unit working.  Now simultaneously
The third step is to get the vertical sampling unit working.  Now simultaneously
apply the 100mV 1MHz squarewave to the signal input of the sampling unit and to
apply the 100 mV, 1 MHz squarewave to the signal input of the sampling unit and to
the external trigger input of the timing unit.  You can use a BNC T-adapter.  It
the external trigger input of the timing unit.  You can use a BNC T-adapter.  It
will corrupt the signal integrity, but at this stage that doesn't matter.  Confirm
will corrupt the signal integrity, but at this stage that doesn't matter.  Confirm
that the strobe pulses are getting to the sampling plug-in.   
that the strobe pulses are getting to the sampling plug-in.   


The fourth step (if the scope is a 567 or 568/230) is to get the digital unit working.
The fourth step (if the scope is a [[567]] or [[568]]/230) is to get the digital unit working.
 
 