Revision as of 03:12, 1 November 2016
The Tektronix SD20 Loopthrough Sampling Head is a single-channel, 20 GHz loop-through sampling head designed for low-loss testing in applications such as microwave systems research and development, digital device characterization and high-speed digital communications circuit design.
The SD20 is nonterminated and keeps losses to a minimum by routing the signal of interest directly through the sampling head without the need of a power divider. The SD20 can also be used for customized TDR measurements of transmission lines and controlled impedance devices. An external signal generator may be used, instead of the SD-24 step generator, to tailor the TDR stimulus to fit a particular situation. For instance, slower slew rates or higher amplitude may be utilized, or you may perform half-sine or impulse testing. In digital communications, the SD20 is useful for viewing and triggering on the clock signal without the use of a power divider.
Key Specifications
Rise time | 17.5 ps |
---|---|
Bandwidth | 20 GHz typ. |
Noise | typically 750 µVRMS, 1.2 m VRMS max. (350 µV typ./550 µV max. with smoothing) |
Max. input | 1.0 Vp-p signal, 1.6 Vp-p (AC+Offset) operating, ±3 V non-destructive |
Sampling rate | 100 Hz - 200 kHz |
Signal connections | 50 Ω SMA connectors |