3S3: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 3: Line 3:
It requires [[P6038]] probes which contain miniature sampling gates.   
It requires [[P6038]] probes which contain miniature sampling gates.   
The risetime is rated at 350 picoseconds.
The risetime is rated at 350 picoseconds.
The sampling gates use four GaAs diodes.  The sampling pulse is
The sampling gates use four GaAs [[sampling diodes]].  The sampling pulse is
produced by the Gate Generator module in the plug-in using a blocking oscillator
produced by the Gate Generator module in the plug-in using a blocking oscillator
driving a step recovery diode and a 6.5cm clipping line.  Balanced pulses are
driving a step recovery diode and a 6.5cm clipping line.  Balanced pulses are

Revision as of 12:23, 17 March 2012

The Tektronix Type 3S3 is a sampling plug-in for 560-series scopes. It requires P6038 probes which contain miniature sampling gates. The risetime is rated at 350 picoseconds. The sampling gates use four GaAs sampling diodes. The sampling pulse is produced by the Gate Generator module in the plug-in using a blocking oscillator driving a step recovery diode and a 6.5cm clipping line. Balanced pulses are sent up the cable to the sampling gate. The output of the sampling gate returns to the plug-in through the same cable, and enters the preamp. The same blocking oscillator that generates the sampling pulses also generates the strobe for the memory circuit. The sampling efficiency is specified to be 2%. The 3S3 is recommended in the Type 111 manual for checking the output pulse risetime.

Tektronix 3S3 Manual