067-0938-00: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
Jaredcabot (talk | contribs) (Added optimized and resized instruction manual) |
||
Line 1: | Line 1: | ||
{{TM500 | mfg=Tektronix | type=067-0938-00 | function=notch filter calibration fixture | class=calibration | image=067-0938-00.jpg | introduced=(?) | discontinued=(?) | designers= | | {{TM500 | ||
manuals= | | mfg=Tektronix | ||
* [ | | type=067-0938-00 | ||
| function=notch filter calibration fixture | |||
| class=calibration | |||
| image=067-0938-00.jpg | |||
| introduced=(?) | |||
| discontinued=(?) | |||
| designers= | |||
| manuals= | |||
* [[Media:070-2868-00.pdf|067-0938-00 Instruction Manual 070-2868-00]] | |||
}} | }} | ||
It is used for measuring the distortion (spectral impurity) of an oscillator by eliminating the fundamental frequency of the oscillator (primarily the [[SG505]]), so the residual harmonics can be measured. | It is used for measuring the distortion (spectral impurity) of an oscillator by eliminating the fundamental frequency of the oscillator (primarily the [[SG505]]), so the residual harmonics can be measured. |
Revision as of 04:04, 13 November 2022
The Tektronix 067-0938-00 is a notch filter calibration fixture plug-in for the TM500 system.
It is used for measuring the distortion (spectral impurity) of an oscillator by eliminating the fundamental frequency of the oscillator (primarily the SG505), so the residual harmonics can be measured.
The circuit topology is a passive double-T RC notch filter with a precision 600 Ω load and a 60 dB attenuator. The module has no connections to the TM500 mainframe, requiring no power supply.
Key Specifications
Notch frequency | 10 Hz, 20 Hz, 100 Hz, 1 kHz, 10 kHz, 20 kHz, 50 kHz, 100 kHz; ±2% |
---|---|
Notch depth | ≥100 dB at 10 Hz and 20 Hz, ≥60 dB otherwise |
Load impedance must be ≥1 MΩ ±5% // ≤75 pF including cabling and spectrum analyzer input capacitance. Input source impedance must be 600 Ω ±5%.