067-0500-00: Difference between revisions
m (clean up) |
No edit summary |
||
Line 4: | Line 4: | ||
|type=067-0500-00 | |type=067-0500-00 | ||
|summary=Calibration fixture | |summary=Calibration fixture | ||
|image=067- | |image=067-500 1a.jpg | ||
|caption=067-0500-00 Front | |caption=067-0500-00 Front | ||
|introduced=(?) | |introduced=(?) | ||
Line 10: | Line 10: | ||
|designers= | |designers= | ||
|manuals= | |manuals= | ||
* [[Media:062-0806-01.pdf|Tektronix 067-0500-00 Manual (-01 | * [[Media:062-0806-01.pdf|Tektronix 067-0500-00 Manual (-01)]] | ||
* [[Media:062-0806-00.pdf|Tektronix 067-0500-00 Manual (-00) ( | * [[Media:062-0806-00.pdf|Tektronix 067-0500-00 Manual (-00)]] (OCR) | ||
}} | }} | ||
The '''Tektronix 067-0500-00''' (also known as [[3M1]]) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in [[560-series scopes]]. | The '''Tektronix 067-0500-00''' (also known as [[3M1]]) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in [[560-series scopes]]. | ||
Line 23: | Line 23: | ||
==Pictures== | ==Pictures== | ||
<gallery> | <gallery> | ||
067-500 1a.jpg | |||
067-500 3.jpg | |||
067-500 5a.jpg | |||
067-0500-00_front.jpg | 067-0500-00_front.jpg | ||
067-0500-00_schem.jpg | 067-0500-00_schem.jpg |
Revision as of 08:20, 5 December 2022
The Tektronix 067-0500-00 (also known as 3M1) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in 560-series scopes.
It consists of a Schmitt trigger that is driven from a single-ended input coming from a BNC connector on the front panel. The Schmitt trigger produces balanced output signals with controlled source impedance. The normalizing capacitor is adjusted using a procedure similar to adjusting probe compensation. When adjusted correctly, the voltage division ratio formed by the plug-in's output impedance and the mainframe's input impedance should be a real value which does not change across the specified frequency range of the scope. After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted to produce clean transient response in the normalized scope.