The Tektronix 176 is a pulsed high-current test fixture for the model 576 curve tracer.
The 176 can be considered a more modern replacement for the 175. Both can supply collector currents up to 200 A.
The 175 uses a rectified mains-frequency transformer secondary current to supply the collector current to the device under test. In contrast, the 176 supplies the collector current by passing stored charge through an SCR. This pulsed technique allows the 176 to measure points far from the origin of the I-V curve while keeping the instrument compact and light and the minimizing thermal effects in the device under test.
The 176 weighs 13 pounds (6 kg), whereas the 175 weighs 84 pounds (38 kg). The 176 has a cooling fan, as does the 175 (for different reasons).
Key Specifications
Collector Sweep Ranges |
|
---|---|
Power Limits | 1000 W, 100 W, and 10 W |
Pulse Width | 300 μs, 80 μs |
Risetime | 80 μs |
Falltime | 10 μs |
Vertical Display Collector Current Range | 0.1 μA/division to 20 A/division in 1-2-5 sequence |
Vertical Display Accuracy 2 A to 20 A | Within 3% of on-screen value |
Max. Base Current Step Generator Output | 20 A |
Pictures
-
-
block diagram
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-