R144 has a [https://en.wikipedia.org/wiki/Permittivity|absolute permittivity]/dielectric constant of ≈2 as a liquid and  near 1 as a gas.
[[File:R114 dielectric constant.pdf|200px|thump|left|]]


[[Category:Instrument repair reports]]
[[Category:Instrument repair reports]]

Revision as of 07:43, 13 July 2023

Possible replacement gas filling

https://groups.io/g/TekScopes/topic/30956868:

HFC236-FA has almost precisely the same characteristics as R114 (boiling point and vapour pressure), is environmentally friendly, and is used both as a refrigerant (a R114 replacement) and in fire extinguishers. So it should be a perfect solution to keeping the P6015 up and running.

R144 has a permittivity/dielectric constant of ≈2 as a liquid and near 1 as a gas.