067-0500-00: Difference between revisions

From TekWiki
Jump to navigation Jump to search
mNo edit summary
m (redirect to 3M1)
Tag: New redirect
Line 1: Line 1:
{{Plugin Sidebar
#REDIRECT [[3M1]]
|manufacturer=Tektronix
|series=560-series scopes
|type=067-0500-00
|summary=calibration fixture
|image=067-500 1a.jpg
|caption=067-0500-00 Front
|introduced=(?)
|discontinued=(?)
|designers=
|manuals=
* [[Media:062-0806-01.pdf|Tektronix 067-0500-00 Manual (-01)]]
* [[Media:062-0806-00.pdf|Tektronix 067-0500-00 Manual (-00)]] (OCR)
}}
The '''Tektronix 067-0500-00''' (also known as [[3M1]]) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in [[560-series scopes]].
 
It consists of a Schmitt trigger that is driven from a single-ended input coming from a BNC connector on the front panel. 
The Schmitt trigger produces balanced output signals with controlled source impedance. 
The normalizing capacitor is adjusted using a procedure similar to adjusting probe compensation. 
When adjusted correctly, the voltage division ratio formed by the plug-in's output impedance and the mainframe's input impedance should be a real value which does not change across the specified frequency range of the scope.
After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted to produce clean transient response in the normalized scope.
 
==Pictures==
<gallery>
067-500 1a.jpg
067-500 3.jpg
067-500 5a.jpg
067-0500-00_front.jpg
067-0500-00_schem.jpg
067-0500-00_insides.jpg
</gallery>
 
[[Category:Calibration fixtures]]
[[Category:560 series plugins]]

Revision as of 09:01, 22 September 2023

Redirect to: