3C66: Difference between revisions
No edit summary |
mNo edit summary |
||
Line 3: | Line 3: | ||
|series=560-series scopes | |series=560-series scopes | ||
|type= 3C66 | |type= 3C66 | ||
|summary=Carrier | |summary=Carrier amplifier plug-in | ||
|image=3c66_front.jpg | |image=3c66_front.jpg | ||
|caption=Tektronix 3C66 | |caption=Tektronix 3C66 | ||
Line 12: | Line 12: | ||
* [[Media:070-357.pdf|Tektronix 3C66 Manual]] | * [[Media:070-357.pdf|Tektronix 3C66 Manual]] | ||
}} | }} | ||
The '''Tektronix 3C66 | The '''Tektronix 3C66''' Carrier Amplifier is a plug-in for [[560-series scopes]], [[introduced in 1963]]. | ||
It is designed to amplify signals from transducers such as strain gauges, whose impedance varies based on, for example, mechanical strain. | It is designed to amplify signals from transducers such as strain gauges, whose impedance varies based on, for example, mechanical strain. | ||
Line 26: | Line 26: | ||
==Pictures== | ==Pictures== | ||
<gallery> | <gallery> | ||
3c66_front.jpg | 3c66_front.jpg |
Revision as of 23:14, 28 November 2023
The Tektronix 3C66 Carrier Amplifier is a plug-in for 560-series scopes, introduced in 1963.
It is designed to amplify signals from transducers such as strain gauges, whose impedance varies based on, for example, mechanical strain. These transducers are usually connected in a bridge circuit.
The 3C66 drives the bridge with a 25 kHz "carrier" signal, which is modulated by the relative impedances of the arms of the bridge. The resulting signal is amplified and envelope-detected by the 3C66, producing a vertical deflection proportional to the measured quantity (strain, for example).
The 3C66 provides built-in variable impedances for balancing the bridge. Because the carrier is 25 kHz, the maximum bandwidth of the measured quantity is 5 kHz. This limitation is due to the settling time of the envelope detector. In addition to the vertical deflection signal, the 3C66 provides a front panel output of the measured signal.
Pictures
-
-
-
-
-
-
-
-
-
-
-
-
-
schematic