P6015: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
(pic, CFC info)
Line 1: Line 1:
The P6015 is a high-voltage probe.  It can be operated with or without its freon filling.  Without the freon filling, the maximum voltage is reduced. The P6015 is a 1000X probe, and it has a compensating circuit
[[File:Tek P6015 box.jpg|thumb|300px|right|Tektronix P6015 in box]]
at the scope end of the cable.
The Tektronix '''P6015''' is a 1000 × high-voltage probe usable up to 20 kV (40 kV peak pulse).  It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak). The original filling material, Freon-114 aka Fluorocarbon 114 (R-114, CFC-114), is no longer available due to its ozone-depleting properties. In later models, a silicone based replacement is used.
 
The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants.
[[File:Tek P6015 derating.jpg|300px|thumb|right|Derating curves]]
{{BeginSpecs}}
{{Spec | Attenuation ratio | 1000 : 1 (variable by 9%) }}
{{Spec | Input impedance | 100 MΩ // 3 pF }}
{{Spec | Rise time | 4.67 ns}}
{{Spec | Bandwidth | 75 MHz }}
{{Spec | Max. input voltage |
* With dielectric filling: 20 kV DC or RMS to 100 kHz, pulse up to 40 kV (10%, < 0.1s)
* Without dielectric filling: 13 kV DC or RMS to 500 kHz, pulse up to 18 kV (10%, < 0.1s)
}}
{{EndSpecs}}
 
==Manuals==


* [http://w140.com/p6015.pdf Tektronix P6015 Manual (Copyright date 1963, PDF)]
* [http://w140.com/p6015.pdf Tektronix P6015 Manual (Copyright date 1963, PDF)]
Line 7: Line 22:
* [http://w140.com/tek_p6015a_mod.pdf Tektronix P6015A Manual with Schematic (PDF)]
* [http://w140.com/tek_p6015a_mod.pdf Tektronix P6015A Manual with Schematic (PDF)]


==Links==
* [http://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_other/P6015.html P6015 @ amplifier.cd]
==Pictures==
<gallery>
<gallery>
File:P6015 BW.png
File:P6015 BW.png

Revision as of 04:04, 26 December 2014

Tektronix P6015 in box

The Tektronix P6015 is a 1000 × high-voltage probe usable up to 20 kV (40 kV peak pulse). It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak). The original filling material, Freon-114 aka Fluorocarbon 114 (R-114, CFC-114), is no longer available due to its ozone-depleting properties. In later models, a silicone based replacement is used.

The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants.

Derating curves

Key Specifications

Attenuation ratio 1000 : 1 (variable by 9%)
Input impedance 100 MΩ // 3 pF
Rise time 4.67 ns
Bandwidth 75 MHz
Max. input voltage
  • With dielectric filling: 20 kV DC or RMS to 100 kHz, pulse up to 40 kV (10%, < 0.1s)
  • Without dielectric filling: 13 kV DC or RMS to 500 kHz, pulse up to 18 kV (10%, < 0.1s)

Manuals

Links

Pictures