The Tektronix 7B92 is a 500 MHz timebase plug-in for [[7000-series scopes]]. It was introduced along with the [[7904]] mainframe in 1972.
The Tektronix 7B92 is a 500 MHz dual timebase plug-in for [[7000-series scopes]].  
It was introduced along with the [[7904]] mainframe in 1972.


The 7B92 uses a discrete trigger circuit containing [[tunnel diodes]].
An "HF Sync" triggering mode is provided in which the trigger level
control varies the frequency of a built-in oscillator to lock on to the input
for  input signals from 100 to 500 MHz, providing higher sensitivity than the
direct trigger (which is specified up to 500 MHz as well).


Up to serial number B069999, the 7B92A used a discrete trigger circuit.
Internal jumpers are provided to configure the 7B92 for operation in 7800/7900 vs. slower mainframes,
After that, the 7B92A used a [[155-0061-00]] trigger amplifier followed by a
and to select whether the variable control affects the delaying or the delayed time base.
[[155-0150-00]] trigger detector.


==Specifications==
==Specifications==
Line 30: Line 33:
{{Spec | Ext Trig input | 1 MΩ // 20 pF or 50 Ω }}
{{Spec | Ext Trig input | 1 MΩ // 20 pF or 50 Ω }}
{{EndSpecs}}
{{EndSpecs}}
==Internals==
The 7B92 uses a discrete trigger circuit containing [[tunnel diodes]].
Up to serial number B069999, the 7B92A used a discrete trigger circuit.
After that, the 7B92A used a [[155-0061-00]] trigger amplifier followed by a
[[155-0150-00]] trigger detector.
==Pictures==
==Pictures==



Revision as of 23:30, 1 October 2014

Template:Plugin Sidebar 2 The Tektronix 7B92 is a 500 MHz dual timebase plug-in for 7000-series scopes. It was introduced along with the 7904 mainframe in 1972.

An "HF Sync" triggering mode is provided in which the trigger level control varies the frequency of a built-in oscillator to lock on to the input for input signals from 100 to 500 MHz, providing higher sensitivity than the direct trigger (which is specified up to 500 MHz as well).

Internal jumpers are provided to configure the 7B92 for operation in 7800/7900 vs. slower mainframes, and to select whether the variable control affects the delaying or the delayed time base.

Specifications

Key Specifications

Sweep speed 0.5 ns/Div to 0.2 s/Div, 1—2—5 sequence (variable up to 0.5 s/Div)
Delay time 0 to 9.9 Div
Triggering 0.5 Div or 100 mV up to 20 MHz, 1 Div or 500 mV up to 600 MHz
Jitter < 50 ps at 600 MHz
Ext Trig input 1 MΩ // 20 pF or 50 Ω

Internals

The 7B92 uses a discrete trigger circuit containing tunnel diodes.

Up to serial number B069999, the 7B92A used a discrete trigger circuit. After that, the 7B92A used a 155-0061-00 trigger amplifier followed by a 155-0150-00 trigger detector.

Pictures