The P6015 is a high-voltage probe. It can be operated with or without | [[File:Tek P6015 box.jpg|thumb|300px|right|Tektronix P6015 in box]] | ||
at the scope end of the cable. | The Tektronix '''P6015''' is a 1000 × high-voltage probe usable up to 20 kV (40 kV peak pulse). It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak). The original filling material, Freon-114 aka Fluorocarbon 114 (R-114, CFC-114), is no longer available due to its ozone-depleting properties. In later models, a silicone based replacement is used. | ||
The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants. | |||
[[File:Tek P6015 derating.jpg|300px|thumb|right|Derating curves]] | |||
{{BeginSpecs}} | |||
{{Spec | Attenuation ratio | 1000 : 1 (variable by 9%) }} | |||
{{Spec | Input impedance | 100 MΩ // 3 pF }} | |||
{{Spec | Rise time | 4.67 ns}} | |||
{{Spec | Bandwidth | 75 MHz }} | |||
{{Spec | Max. input voltage | | |||
* With dielectric filling: 20 kV DC or RMS to 100 kHz, pulse up to 40 kV (10%, < 0.1s) | |||
* Without dielectric filling: 13 kV DC or RMS to 500 kHz, pulse up to 18 kV (10%, < 0.1s) | |||
}} | |||
{{EndSpecs}} | |||
==Manuals== | |||
* [http://w140.com/p6015.pdf Tektronix P6015 Manual (Copyright date 1963, PDF)] | * [http://w140.com/p6015.pdf Tektronix P6015 Manual (Copyright date 1963, PDF)] | ||
Line 7: | Line 22: | ||
* [http://w140.com/tek_p6015a_mod.pdf Tektronix P6015A Manual with Schematic (PDF)] | * [http://w140.com/tek_p6015a_mod.pdf Tektronix P6015A Manual with Schematic (PDF)] | ||
==Links== | |||
* [http://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_other/P6015.html P6015 @ amplifier.cd] | |||
==Pictures== | |||
<gallery> | <gallery> | ||
File:P6015 BW.png | File:P6015 BW.png |
Revision as of 04:04, 26 December 2014
The Tektronix P6015 is a 1000 × high-voltage probe usable up to 20 kV (40 kV peak pulse). It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak). The original filling material, Freon-114 aka Fluorocarbon 114 (R-114, CFC-114), is no longer available due to its ozone-depleting properties. In later models, a silicone based replacement is used.
The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants.
Key Specifications
Attenuation ratio | 1000 : 1 (variable by 9%) |
---|---|
Input impedance | 100 MΩ // 3 pF |
Rise time | 4.67 ns |
Bandwidth | 75 MHz |
Max. input voltage |
|
Manuals
- Tektronix P6015 Manual (Copyright date 1963, PDF)
- Tektronix P6015 Manual (Revision date 1985, PDF)
- Tektronix P6015A Manual (PDF)
- Tektronix P6015A Manual with Schematic (PDF)