* [http://amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/calibration-fixture.htm 067-0587-02 page @ amplifier.cd] (comparing -01 and - | * [http://amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/calibration-fixture.htm 067-0587-02 page @ amplifier.cd] (comparing -01 and -02, lots of pictures, text in German) | ||
==Pictures== | ==Pictures== |
Revision as of 11:57, 22 January 2015
Template:Plugin Sidebar 2 The Tektronix 067-0587-02 is a signal standardizer (calibration) plug-in for 7000-series scopes. It is a complete redesign of the 067-0587-01 standardizer. It provides a calibrated staircase at 1 line/Div in Gain mode, a fast-rising pulse in Step Resp mode,
Key Specifications
— GAIN mode — | |
Staircase accuracy | ± 0.3% in ± 3 Div, ± 0.5% in ± 4 Div, Linearity ± 0.2% of 10 Div |
— STEP RESP mode — | |
Amplitude | 2 to 10 Div |
Position | Top at least ±5 Div from center of screen |
Rise time | < 150 ps |
Repetition rate | 10 Hz to 1 MHz in decade steps, 0.1 % |
Pretrig Out | 0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable) |
— FREQ RESP mode — | |
Input impedance | 50 Ω |
Input range | 0.5 to 1 Vp-p |
Flatness | ± 3% from 3 MHz to 1 GHz |
Internals
The 067-0587-02 is based on a number of Tektronix custom ICs from the 1 GHz 7104 scope generation, such as the 155-0160-00 trigger amplifier + source select chip and the 155-0175-00 trigger amplifier chip.
The mode selector switch is a special elastomer-based cam switch. The manual warns that while it can be cleaned if proper procedure is followed (note some screws are labelled "do not remove" on the cover!), only at the factory can (could ...) it be repaired/re-assembled completely.
Links
- 067-0587-02 page @ amplifier.cd (comparing -01 and -02, lots of pictures, text in German)