The Tektronix DM501 is a 4½ digit True RMS multimeter plug-in for the TM500 system.
It also has a temperature input, requiring the P6058 temperature probe (DM501A: P6601 PT-100 probe).
Key Specifications
DC V ranges | 200 mV, 2 V, 20 V, 200 V, 1000 V |
---|---|
DC V accuracy | 0.05% of reading + 0.01% FS in 2 V to 200 V ranges @ +18 °C to +28 °C |
DC V input resistance | 10 MΩ |
AC V ranges | 200 mV, 2 V, 20 V, 200 V, 500 V |
AC V accuracy | 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C |
AC V input resistance | 10 MΩ // 160 pF |
dB ranges | -40, -20, 0, +20, +40 dB |
Ω ranges | 200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20 MΩ |
Ω accuracy | 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C |
DC A ranges | 200 μA, 2 mA, 20 mA, 200 mA, 2 A |
DC A accuracy | 0.2% of reading + 0.015% FS @ +18 °C to +28 °C |
AC A ranges | 200 μA, 2 mA, 20 mA, 200 mA, 2 A |
AC A accuracy | 0.6% of reading + 0.05% FS, 20 Hz to 10 kHz @ +18 °C to +28 °C |
Temperature range | -62 °C to +240 °C |
Temperature accuracy | ±2 °C from -62 °C to +150 °C with probe calibrated to instrument, @ +18 °C to +28 °C ambient |
Pictures
-
DM501
-
DM501A
-
DM501A