{{BeginSpecs}} | {{BeginSpecs}} | ||
{{Spec | Deflection | 10 μV/Div to 10 V/Div in 1–2–5 sequence }} | {{Spec | Deflection | 10 μV/Div to 10 V/Div in 1–2–5 sequence }} | ||
{{Spec | Input impedance | 1 MΩ // 47 pF ( | {{Spec | Input impedance | 1 MΩ // 47 pF (Volts mode) or 10 MΩ (transducer mode) }} | ||
{{Spec | Bandwidth | 1 MHz, LF limit switchable DC, 0.1 Hz to 10 kHz in six steps, HF limit switchable 100 Hz to 1 MHz in ×3/×10 steps, AC coupling 1.6 Hz }} | {{Spec | Bandwidth | 1 MHz, LF limit switchable DC, 0.1 Hz to 10 kHz in six steps, HF limit switchable 100 Hz to 1 MHz in ×3/×10 steps, AC coupling 1.6 Hz }} | ||
{{Spec | Signal ranges | }} | {{Spec | Signal ranges | }} |
Revision as of 05:00, 20 November 2018
Template:Plugin Sidebar 2 The Tektronix 3A10 is a differential amplifier plug-in for 560-series scopes.
It is intended for use with transducers and for that purpose it has a DC-coupled differential input and a variable voltage supply for powering the transducer. It also has a 6-pin input connector (Bendix MS3101A-10-6S) to directly connect to the 015-0169-00 Strain Gage Adapter.
The 3A10 circuit is partially based on the 1A7A's.
Key Specifications
Deflection | 10 μV/Div to 10 V/Div in 1–2–5 sequence |
---|---|
Input impedance | 1 MΩ // 47 pF (Volts mode) or 10 MΩ (transducer mode) |
Bandwidth | 1 MHz, LF limit switchable DC, 0.1 Hz to 10 kHz in six steps, HF limit switchable 100 Hz to 1 MHz in ×3/×10 steps, AC coupling 1.6 Hz |
Signal ranges |
Range | Differential signal | DC Offset | Common mode |
---|---|---|---|
10 μV/Div to 10 mV/Div | ±1 V | ±1 V | ±10 V |
20 mV/Div to 0.1 V/Div | ±10 V | ±10 V | ±100 V |
0.2 V/Div to 1 V/Div | ±100 V | ±100 V | ±500 V |
2 V/Div to 10 V/Div | ±1000 V | ±1000 V | ±500 V |
Pictures
-
-
-
-
-
top
-
right
-
left
-
front
-
bottom
-
front