Nelson-Ross PSA-011: Difference between revisions
(subcat) |
No edit summary |
||
Line 8: | Line 8: | ||
series=[[500-series scopes]]| | series=[[500-series scopes]]| | ||
manuals= | manuals= | ||
* [ | * [[Media:Nelson-ross psa-001 002 003 011 012 013 plug-ins instruction sch.pdf|Manual for Nelson-Ross 001, 002, 003, 011, 012, and 013 (PDF)]] | ||
}} | }} | ||
The '''Nelson-Ross PSA-011''' is a spectrum analyzer plug-in for [[500-series scopes]]. | The '''Nelson-Ross PSA-011''' is a spectrum analyzer plug-in for [[500-series scopes]]. |
Revision as of 06:01, 29 April 2019
Template:Plugin Sidebar 2 The Nelson-Ross PSA-011 is a spectrum analyzer plug-in for 500-series scopes. The center frequency range is 10 Hz to 20 kHz. It can display amplitude in linear or log scale.
The input is high-impedance and is AC-coupled. The PSA-011 is a single-conversion architecture with a pair of 6BH6 tubes serving as a balanced mixer. After IF filtering and amplification, the signal is detected using a 1N34A germanium diode. The detected voltage from the diode directly drives pin 3 (vertical -) of the scope plug-in interface connector. Pin 1 (vertical +) of the scope plug-in interface connector is fed a DC vertical position voltage by the PSA-011.
The voltage-controlled local oscillator is tuned using a variable-reactance tube, V6, a 6201.
There were at least two major versions produced, without changing the model number. Compare the picture from the manual below with the other front panel photographs.
There was also a tracking generator, called the Nelson Ross 101 SYNCHRO-SWEEP Generator. These are very rare.