2336: Difference between revisions

From TekWiki
Jump to navigation Jump to search
mNo edit summary
No edit summary
Line 20: Line 20:


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Bandwidth and Channels| 100 MHz in a 1, 2, 5 sequence; Bandwidth is at least 88 MHz at 2 mV/div; 2 Channels }}
{{Spec | Bandwidth | 100 MHz (≥88 MHz at 2 mV/div) }}
{{Spec | Deflection Factor| CH 1, 2: 5 mV/div to 5 V/div, all ±3%; Variable control can increase sensitivity to at least 2 mV/div.}}
{{Spec | Deflection | 5 mV/div to 5 V/div (1−2−5 sequence), ±3%; Variable control can increase sensitivity to at least 2 mV/div}}
{{Spec | Input Impedance| 1 MΩ ±2% and 20 pF ±10% }}
{{Spec | Input Impedance | 1 MΩ ±2% // 20 pF ±10%, 2 channels }}
{{Spec | Maximum Input Voltage| 400 V (DC + peak AC) or 500 V (p-p ac at ≤ 1 kHz) }}
{{Spec | Maximum Input Voltage| 400 V (DC + peak AC) or 500 V (p-p ac at ≤ 1 kHz) }}
{{Spec | Sweep Speeds| A Time Base: 0.5 s/div to 50ns/div; B Time Base: 50 ms/div to 50 ns/div; A and B X10 magnifier extends max sweep to 5 ns/div }}
{{Spec | Sweep | A Time Base: 0.5 s/div to 50 ns/div, 1−2−5; B Time Base: 50 ms/div to 50 ns/div, 1−2−5; A and B ×10 magnifier extends max sweep to 5 ns/div }}
{{Spec | Trigger Source| VERT MODE, CH 1, CH 2, LINE, EXT, and EXT ÷ 10 }}
{{Spec | Trigger Source| VERT MODE, CH 1, CH 2, LINE, EXT, and EXT ÷ 10 }}
{{Spec | Trigger Coupling| AC, LF REJ, HF REJ, or DC }}
{{Spec | Trigger Coupling| AC, LF REJ, HF REJ, or DC }}
{{Spec | Probes| Two P6108A 100 MHz X10 probes, BNC connector with a readout pin }}
{{Spec | Probes| Two [[P6108A]] 100 MHz ×10 probes, BNC connector with a readout pin }}
{{Spec | Power| 40 W max, 90 − 250 VAC, 48 − 440 Hz }}
{{Spec | Power| 40 W max, 90 − 250 V<sub>AC</sub>, 48 − 440 Hz }}
{{EndSpecs}}
{{EndSpecs}}


Line 37: Line 37:


<gallery>
<gallery>
Tektronix_2336.jpg
Tektronix 2336.jpg
</gallery>
</gallery>


  [[Category:2300 series scopes]]
  [[Category:2300 series scopes]]

Revision as of 03:56, 4 November 2020

{{{manufacturer}}} 
Ruggedized 100 MHz dual trace scope
Tektronix 2336

Produced from 1982 to 1991

Manuals
Manuals – Specifications – Links – Pictures


The Tektronix 2336 is a ruggedized portable dual-channel, 100 MHz analog scope. An A Ext connector (external signal to the A Trigger Generator) on the front panel, B Ext connector (external signal to the B Trigger Generator) on the front cover, and External Z-Axis connector at the rear round out its connectivity. The front cover also has a 3 ½ digit LCD for delta TIME measurements and B Trigger function. A 150 MHz modification was available.

Vertical controls were by way of calibrated deflection factors on CH 1 or CH 2 inputs through AC, DC, or GND coupling. The vertical mode selections are CH 1, ALT, CHOP (~275 kHz), ADD, CH 2, AUTO, X-Y, TRIG VIEW, or BW LIMIT (20 MHz). Horizontal deflection is controlled by either the calibrated A or B Sweep Generators, or Alternate between the two of the A, and B Delayed Sweep.

Trigger sources may be VERT MODE, CH 1, CH 2, LINE, EXT, and EXT ÷ 10. Trigger coupling can be AC, LF REJ, HF REJ, or DC, and can be controlled by amplitude level and slope type. The Trigger Modes may be set to AUTO, NORM, and SGL SWP.

Key Specifications

Bandwidth 100 MHz (≥88 MHz at 2 mV/div)
Deflection 5 mV/div to 5 V/div (1−2−5 sequence), ±3%; Variable control can increase sensitivity to at least 2 mV/div
Input Impedance 1 MΩ ±2% // 20 pF ±10%, 2 channels
Maximum Input Voltage 400 V (DC + peak AC) or 500 V (p-p ac at ≤ 1 kHz)
Sweep A Time Base: 0.5 s/div to 50 ns/div, 1−2−5; B Time Base: 50 ms/div to 50 ns/div, 1−2−5; A and B ×10 magnifier extends max sweep to 5 ns/div
Trigger Source VERT MODE, CH 1, CH 2, LINE, EXT, and EXT ÷ 10
Trigger Coupling AC, LF REJ, HF REJ, or DC
Probes Two P6108A 100 MHz ×10 probes, BNC connector with a readout pin
Power 40 W max, 90 − 250 VAC, 48 − 440 Hz