067-0587-10: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 20: Line 20:
generates a pre-trigger signal on the PRETRIG OUT connector while in any mode.
generates a pre-trigger signal on the PRETRIG OUT connector while in any mode.


{{MissingSpecs}}
{{BeginSpecs}}
{{SpecGroup | GAIN mode }}
{{Spec | Staircase accuracy | ± 0.3% in ± 3 Div, ± 0.5% in ± 4 Div, Linearity ± 0.2% of 10 Div }}
{{SpecGroup |  STEP RESP mode }}
{{Spec | Amplitude | 2 to 10 Div }}
{{Spec | Position  | Top at least ±5 Div from center of screen }}
{{Spec | Rise time | < 150 ps}}
{{Spec | Repetition rate | 10 Hz to 1 MHz in decade steps, 0.1 %}}
{{Spec | Pretrig Out | 0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable) }}
{{SpecGroup |  FREQ RESP mode }}
{{Spec | Input impedance | 50 Ω }}
{{Spec | Input range    | 0.5 to 1 V<sub>p-p</sub> }}
{{Spec | Flatness | ± 3% from 3 MHz to 1 GHz }}
{{EndSpecs}}


==Pictures==
==Pictures==

Revision as of 20:20, 7 March 2021

Template:Plugin Sidebar 2 The Tektronix 067-0587-10 is a Calibration fixture / signal standardizer for 7000-series scopes. It has the same performance specifications as the 067-0587-02 but is modified for compatibility with the 7612D.

In the 067-0587-10, the signal coming from the PRETRIG OUT generates one trigger for each staircase signal when used with the TEST switch in the GAIN position. The 067-0587-02 generates a pre-trigger signal on the PRETRIG OUT connector while in any mode.

Key Specifications

     — GAIN mode
Staircase accuracy ± 0.3% in ± 3 Div, ± 0.5% in ± 4 Div, Linearity ± 0.2% of 10 Div
     — STEP RESP mode
Amplitude 2 to 10 Div
Position Top at least ±5 Div from center of screen
Rise time < 150 ps
Repetition rate 10 Hz to 1 MHz in decade steps, 0.1 %
Pretrig Out 0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable)
     — FREQ RESP mode
Input impedance 50 Ω
Input range 0.5 to 1 Vp-p
Flatness ± 3% from 3 MHz to 1 GHz

Pictures