The '''Tektronix 11A81''' is a 3 GHz single-channel sampler/amplifier plug-in for [[11000-series scopes]], primarily the [[11403A]].
The '''Tektronix 11A81''' is a 3 GHz single-channel sampler/amplifier plug-in for the [[11403A]] and the [[CSA404]].
The 11A81 receives an "acquisition strobe" signal from the mainframe, and samples the input signal.
The 11A81 receives an "acquisition strobe" signal from the mainframe, and samples the input signal.
The sampled signal is fed to the mainframe, where it is digitized and displayed.
The sampled signal is fed to the mainframe, where it is digitized and displayed.

Revision as of 23:21, 7 March 2021

Template:Plugin Sidebar 2 The Tektronix 11A81 is a 3 GHz single-channel sampler/amplifier plug-in for the 11403A and the CSA404. The 11A81 receives an "acquisition strobe" signal from the mainframe, and samples the input signal. The sampled signal is fed to the mainframe, where it is digitized and displayed. This architecture allows the 11A81 to provide higher bandwidth than the 11k plug-in to mainframe signal path.

Development of the 11A81 was managed by Murlan Kaufman.

A 165-2448-00 (U950) diode bridge IC is used for over-voltage protection. A 165-2392-00 (U650) H2392 sampling bridge performs the sampling.

Key Specifications

Bandwidth DC to 3 GHz (in 11403A/CSA404 only)
Rise time < 130 ps
Deflection 10 mV to 1V per division
Input impedance 50 Ω

Pictures