Line 1: | Line 1: | ||
__TOC__ | __TOC__ | ||
==Disintegrated fan mounts== | |||
: See [[Fan mounts]]. | |||
==Supply out of regulation== | ==Supply out of regulation== | ||
:'''Symptom:''' +350 V supply out of regulation (~300 V) | :'''Symptom:''' +350 V supply out of regulation (~300 V) |
Latest revision as of 10:58, 6 July 2022
Disintegrated fan mounts
- See Fan mounts.
Supply out of regulation
- Symptom: +350 V supply out of regulation (~300 V)
- Root cause: C751 10 nF "Bumble Bee" leaky
- Solution: Replace C751
HV very low
- Symptom: HV very low, oscillator tube V800 gets very hot
- Root cause: Black Beauty capacitors in HV circuit leaky
- Solution: Replace all 6.8 nF HV caps
No sweep at slower speeds
- Symptom: In 1 ms/Div and slower, no sweep − trace stationary on left
- Root cause: Bumble Bee capacitors in holdoff circuit leaky
- Solution: Replace C180B-D
Burned-out HV transformer
- Symptom: HV transformer showing burns
- Root cause: overload by earlier V800 failure caused by leaky HV caps
- Solution: Replaced by 120-114 transformer (502) with 120-150 (from 502A) using only one cathode secondary. Note the grid winding has more turns on the 120-150, so a dropper resistor needs to be added to prevent the grid going too negative (no beam).