P6015: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(first in 1963 catalog)
No edit summary
Line 9: Line 9:
|discontinued=(?)
|discontinued=(?)
|manuals=
|manuals=
* [[Media:070-0373-01.pdf|P6015 Manual 070-0373-01]] (Copyright date 1963)
* [[Media:070-0373-01.pdf|P6015 Manual 070-0373-01]] (C 1963)
* [[Media:070-0373-02.pdf|P6015 Manual 070-0373-02]] (Revision date 1985)
* [[Media:070-0373-02.pdf|P6015 Manual 070-0373-02]] (Rev. 1985)
* [[Media:070-0373-02 rev 1990.pdf|P6015 Manual 070-0373-02]] (Rev. 1990)
* [[Media:070-8223-02.pdf|P6015A Manual 070-8223-02]]
* [[Media:070-8223-02.pdf|P6015A Manual 070-8223-02]]
* [[Media:070-8223-04.pdf|P6015A Manual 070-8223-04]] (with schematics)
* [[Media:070-8223-04.pdf|P6015A Manual 070-8223-04]] (with schematics)

Revision as of 00:16, 19 July 2022

Tektronix P6015
20(40) kV, ×1000 Passive Probe
P6015

Compatible with 1 MΩ scope inputs

Produced from 1963 to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6015 is a 1000× high-voltage probe usable up to 20 kV (40 kV peak pulse).

It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak), similar to the P6013 predecessor model.
The original filling material, Freon-114 aka Fluorocarbon 114 (R-114, CFC-114), is no longer available due to its ozone-depleting properties.

In the CFC-free P6015A successor, introduced in 1991, a silicone based replacement is used.

The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants.

Key Specifications

Bandwidth 75 MHz (10' cable), 25 MHz (25' cable)
Risetime 4.67 ns (10' cable), 14 ns (25' cable)
Attenuation 1000×
Input impedance 100 MΩ // 3 pF
Compensation range 15-50 pF
Max. input voltage
  • With dielectric filling: 20 kV DC or RMS to 100 kHz, pulse up to 40 kV (10%, < 0.1s)
  • Without dielectric filling: 13 kV DC or RMS to 500 kHz, pulse up to 18 kV (10%, < 0.1s)

Links

Pictures