The Tektronix P6129 is a passive 100 MHz ×1/×10 miniature modular probe, with a BNC connector that has a readout pin.
Key Specifications
Attenuation | ×1/×10 ±1.35% at DC |
---|---|
Input Impedance | ×1: 1 MΩ ±1.0% // 90–115 pF; ×10: 10 MΩ ±1.3% // 18.5 pF (17.5–19.5 pF) |
Compensation Range | 15–35 pF |
Bandwidth | ×1: DC–10 MHz; ×10: DC–100 MHz |
Maximum Input Voltage | 420 V (DC + peak AC) nondestructive |