The Tektronix 491 is a spectrum analyzer with a frequency range of 10 MHz to 12.4 GHz (up to 40 GHz with external waveguide mixers). The 491 was introduced in August 1966. Due to severe performance limitations, the 491 was replaced by the completely redesigned 492 in 1980.
Key Specifications
Frequency Range |
|
---|---|
Dispersion Range | (plus 0 Hz) 200 kHz/div to 10 MHz/div in 1−2−5 sequence |
Resolution Bandwidth | 1 kHz to 100 kHz (6 dB) |
Input impedance | 50 Ω |
Input Power | max. −30 dBm for linear operation, +15 dBm (25 mW) diode mixer power limit |
IF Attenuator Range | 0 dB to 51 dB, in combinations of 1, 2, 4, 8, 16, and 20 dB |
Sweep Speed Range | 0.5 s/div to 10 µs/div in 1, 2, 5 sequence; Uncalibrated to approx 1.25 s/div |
Triggering Modes | Internals, External, and Line; 100 V maximum input (DC + peak AC) |
Dynamic Range | 80 dB (Max sensitivity of −110 dBm minus Max input power of −30 dBm); display range ≥40 dB |
Calibrator Frequency | 200 MHz |
Weight | 13.6 kg / 30 lbs (Rackmount version 18 kg / 40 lbs) |
Power | 90−136 and 180−272 VAC in six ranges, 48−440 Hz. At 115 VAC, 60 Hz, 55 W |
Front panel inputs/outputs |
|
Rear panel inputs/outputs |
|
Options
- Opt. 01: Limit Frequency Range to 10 MHz to 2 GHz
- Opt. 02: Limit Frequency Range to 1.5 GHz to 40 GHz
Prices
1979: $10,700 ($41,800 in 2022 Dollars)
According to an internal memo, in 1979 annual sales were estimated at 172 units.