P6010: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
m (No mention of use for sampling inputs in the P6010 manual.) |
||
Line 12: | Line 12: | ||
* [[Media:070-0495-01.pdf|Tektronix P6010 Manual]] (OCR) | * [[Media:070-0495-01.pdf|Tektronix P6010 Manual]] (OCR) | ||
}} | }} | ||
The '''Tektronix P6010''' is a | |||
The ''' Tektronix P6010''' is a x10 miniature passive probe designed for use with Tektronix wide-band oscilloscopes. | |||
{{BeginSpecs}} | {{BeginSpecs}} |
Latest revision as of 07:32, 14 February 2023
The Tektronix P6010 is a x10 miniature passive probe designed for use with Tektronix wide-band oscilloscopes.
Key Specifications
Attenuation | ×10 |
---|---|
Rise time | ≤2 ns (probe), 7 ns with 453 scope |
Bandwidth | 150 MHz |
Input impedance | 10 MΩ // <10–14.5 pF, depending on cable length |
Compensation range | 14-21 pF, factory compensated to 20 pF |
Maximum voltage | 500 V DC + peak AC, derate above 12 MHz |
Cable length | 3.5, 6, or 9 feet long, terminated in a BNC connector |