224/Repairs: Difference between revisions
Jump to navigation
Jump to search
(Created page with "On the 224 and 222 scopes, the CRT neck socket is known to come loose or break. The clips that are meant to hold it together become brittle and break off easily, requiring sup...") |
No edit summary |
||
Line 1: | Line 1: | ||
On the 224 and 222 scopes, the CRT neck socket is known to come loose or break. The clips that are meant to hold it together become brittle and break off easily, requiring super glue to fix it. A loose socket would also be a reason for in appropriate blanking of the trace and its vector graphics. | ==CRT socket issue== | ||
On the 224 and 222 scopes, the CRT neck socket is known to come loose or break. The clips that are meant to hold it together become brittle and break off easily, requiring super glue to fix it. | |||
A loose socket would also be a reason for in appropriate blanking of the trace and its vector graphics. | |||
[[Category:Instrument repair reports]] |
Revision as of 06:16, 25 July 2023
CRT socket issue
On the 224 and 222 scopes, the CRT neck socket is known to come loose or break. The clips that are meant to hold it together become brittle and break off easily, requiring super glue to fix it.
A loose socket would also be a reason for in appropriate blanking of the trace and its vector graphics.