224/Repairs: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
Line 2: Line 2:
On the 224 and 222 scopes, the CRT neck socket is known to come loose or break. The clips that are meant to hold it together become brittle and break off easily, requiring super glue to fix it.
On the 224 and 222 scopes, the CRT neck socket is known to come loose or break. The clips that are meant to hold it together become brittle and break off easily, requiring super glue to fix it.


A loose socket would also be a reason for in appropriate blanking of the trace and its vector graphics.
A loose socket would also be a reason for inappropriate blanking of the trace and its vector graphics.


[[Category:Instrument repair reports]]
[[Category:Instrument repair reports]]

Latest revision as of 10:33, 27 July 2023

CRT socket issue

On the 224 and 222 scopes, the CRT neck socket is known to come loose or break. The clips that are meant to hold it together become brittle and break off easily, requiring super glue to fix it.

A loose socket would also be a reason for inappropriate blanking of the trace and its vector graphics.