Bad TI IC sockets: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
Line 44: Line 44:


==Links==
==Links==
* [[Media:Tek_component_news_num_273_aug_24_1979.pdf|Tek Component News No.273 August 1979 – Page 7 – IC socket problem update]]
* [[Media:Component news 273.pdf|Tek Component News No.273, August 1979 – Page 7 – IC socket problem update]]
* [[Media:Component news 261.pdf|Tek Component News No.261, August 1978 – Page 4 – Low-profile IC socket modified]]:
<blockquote>
Recently, the C93 series of low -profile sockets (manufactured by Texas Instruments) was replaced by the C95 series.
This was done to alleviate flux entrapment problems, common in the C93 series.
The newer, C95 sockets have no anti-wicking wafers (see illustrations) , but have demonstrated other inadequacies.
 
C95 series sockets are extremely susceptible to damage when inserting an IC.
Therefore, we were finding many parts with broached entries and/or collapsed contacts.
 
In this case, the cure was worse than the cold !
</blockquote>


==Pictures==
==Pictures==

Latest revision as of 13:07, 17 May 2024

Texas Instruments made some bad IC sockets back in late 1970's. The TI C-95 model of socket is known to be a problem. A typical high quality replacement would be a Mill-Max machined-pin DIP socket.

DIP sockets are often problematic as instruments age. Gold-plated pins carrying low current in gold-plated sockets are the least problematic. Higher currents, such as occur in ECL logic, are more problematic.

For gold pin DIPs like many of the Tektronix custom ICs, one would choose a gold internal contact plating, and for other early non RoHS lead-tin plated ICs a similar internal contact plating is available from some suppliers.

Tek issued a service note (see below) that describes a transition from the original C93 socket design to a C95 design which was mechanically too fragile, and then a transition to an improved version of the C93 socket design.

For the 14 pin DIP at the left, a gold part number would be Mill-Max 110-93-314-41-001000, a non-RoHS lead-tin part number would be Mill-Max 110-99-314-41-001000. Both parts are non-RoHS since the solder-tail (PCB end) is also lead-tin plated.

Former Tek Bench Technician and Applications Engineer Jim Mauck says:

I noticed your article on bad Texas Instruments IC sockets. I literally replaced 100's of those sockets. I believe TI's part number was C95. They were used throughout Tek at the time but they do not always cause an issue. They didn't have a gas-tight contact with the IC pin and they would oxidize over time and develop a resistive connection. This wasn't necessarily a problem for low current applications but it was death to any ECL circuits. Unfortunately ECL was used extensively in logic analyzers. There were very high failure rates for the 7D01, DF1 and DF2, and eventually Tek had an unpublicized board exchange program for those instruments. Service technicians could replace the boards but there was no official modification kit or customer notification.

Failures due to these sockets were typically intermittent. A classic indicator of bad sockets was having the problem disappear if you wiggled all of the ICs in the socket. Doing that scraped off some of the oxidation and the instrument would work for another month or so until the contacts were again oxidized. The problem was exacerbated in some cases because Tek was using ICs with tinned leads intended to be soldered to the circuit board - not socketed. I never saw any problems in high current applications when the leads were gold plated.

Links

Recently, the C93 series of low -profile sockets (manufactured by Texas Instruments) was replaced by the C95 series. This was done to alleviate flux entrapment problems, common in the C93 series. The newer, C95 sockets have no anti-wicking wafers (see illustrations) , but have demonstrated other inadequacies.

C95 series sockets are extremely susceptible to damage when inserting an IC. Therefore, we were finding many parts with broached entries and/or collapsed contacts.

In this case, the cure was worse than the cold !

Pictures