C30 and C32 (100 uF, 25V) have a tendency to leak or short, causing associated resistors R30 and R32 (240, .25W, 5%) to overheat and fail. A good preventative maintenance procedure would be to replace these two capacitors with modern equivalents. Two 5CT1N units purchased online had this fault, and replacement of the capacitors and resistors restored proper operation. | C30 and C32 (100 uF, 25V) have a tendency to leak or short, causing associated resistors R30 and R32 (240, .25W, 5%) to overheat and fail. A good preventative maintenance procedure would be to replace these two capacitors with modern equivalents. Two 5CT1N units purchased online had this fault, and replacement of the capacitors and resistors restored proper operation. | ||
[[Category:Curve tracers]] | |||
[[Category:Specifications needed]] |
Revision as of 07:26, 9 May 2014
Curve Tracer Plug-in for the 5000 series oscilloscopes.
Common Problems
C30 and C32 (100 uF, 25V) have a tendency to leak or short, causing associated resistors R30 and R32 (240, .25W, 5%) to overheat and fail. A good preventative maintenance procedure would be to replace these two capacitors with modern equivalents. Two 5CT1N units purchased online had this fault, and replacement of the capacitors and resistors restored proper operation.