067-0657-00

From TekWiki
Revision as of 00:07, 10 March 2023 by Peter (talk | contribs)
Jump to navigation Jump to search
Tektronix 067-0657-00
Normalized ramp generator
067-0657-00 front

Compatible with 7000-series scopes

Produced from 1976 to

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 067-0657-00 Normalized Ramp Generator is a calibration plug-in for 7000-series scopes.

It provides calibrated sweep rates from 0.5 ns/Div to 50 ns/Div (in a 1–2–5 sequence) plus 1 μs/Div and 1 ms/Div.

The calibration fixture is triggered from an external signal (such as a trigger output from a time-mark generator) applied to the front-panel TRIG IN connector.

Internals

The 067-0657-00 was the result of some fundamental research by Bruce Hofer inspired by production issues in the calibration of early 7904 scopes. This lead to a new sweep circuit that uses current steering in a differential pair to start the charging of the timing capacitor as is discussed in patent 4009399. The sweep circuit in the 067-0657-00 follows the patent application but does not implement the baseline stability circuit, which was incorporated in the 7B92A.

Links

Pictures