Telequipment Oscilloscope Calibrator C3

Revision as of 13:47, 20 February 2022 by Peter (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

The Telequipment C3 is a solid state oscilloscope calibrator.

Telequipment Oscilloscope Calibrator C3
Oscilloscope Calibrator

Produced from (?) to (?)

Manuals

Please add

(All manuals in PDF format unless noted otherwise)

From the manual:

The Calibration Unit Type C3 is a small portable mains-operated instrument for use in service centres.

It provides a wide range of output signals, with a high degree of accuracy as shown in the Specification, for test and calibration purposes.

Key Specifications

Markers 100 ns to 5 s, 1–2–5; pulse amplitude ~1 V; pulse width ~2% of period; accuracy ±0.01%
Amplitude Calibrator 10 kHz / 1 kHz / 50 Hz / 10 Hz / 1 Hz square wave, or DC; amplitude 100 V to 0.025 V and 100 mV to 0.025 mV, 1–2–5; rise time ≤ 100 ns; 50 Hz output has rise time of 400 μs. Accuracy ±0.2% when loaded with 1 MΩ except 0.05 V/mV ±0.3% and 0.025 V/mV ±0.35%; output impedance varies from 5 kΩ at 50 V and 50 mV down to 75 Ω at ≤0.5 V and 0.5 mV.
Fast Rise Signal Square waveform switched at 100 kHz, 1 MHz, 10 MHz; rise time ≤ 1 ns; output 0.2 V max. into 50 Ω
Television Waveform 625-line signal, +ve and —ve modulation; amplitude 1.6 Vp-p; output impedance 75 Ω; -ve going output is DC coupled from ground, +ve going output is AC coupled
TV signal modulation Black, White, or square wave at 62.5 kHz / 250 kHz / 1 MHz / 5 MHz; Modulation can be switched on and off at one-second rate.
Auxiliary Outputs Available on 4 mm sockets at the side; square waves at 10 MHz, 1 MHz, 100 kHz, 10 kHz, 1 kHz, 100 Hz, 10 Hz, 1 Hz; amplitude ~2.5 V, output impedance ~1 kΩ
Ext Clock External clock frequencies of 2 V peak to peak amplitude may be fed to the 10 MHz Auxiliary Output socket
Power 100, 110, 120, 200, 220, 240 V, 48 to 440 Hz, 13 VA.
Size 107 mm × 228 mm × 285 mm (H×W×D)