067-0587-01: Difference between revisions

1,031 bytes added ,  16 October 2014
internals
(cat)
(internals)
Line 1: Line 1:
The 067-0587-01 Calibration Fixture is designed to standardize the
horizontal and vertical deflection factors of 7000 series mainframes.
It provides a uniform step calibrated deflection factor, a fast risetime
step response function and a common-mode input test for both the
vertical and trigger systems.
It uses one custom Tektronix IC, the [[155-0078-00]].  The manual says it
is a high frequency push-pull amplifier.
It uses one tunnel diode, the [[152-0177-00]], a 10 mA, 4 pF diode in
a DO-17 package.  It also uses 12 Tektronix custom Schottky diodes, part [[152-0442-00]]
(speced as matched pairs) in the tunnel diode pulse shaper and in the CW peak-to-peak
detector.  The service manual
calls them "diconnect diodes", physically they look like snap-off diodes
found in other Tektronix fast risetime pulsers.  Note that there is a typo
in the service manual for CR80, CR84, and CR85 in the peak-to-peak detector.
==Specifications==
{{BeginSpecs}}
{{Spec | Risetime | 300 ps}}
{{Spec | Bandwidth | 500 MHz}}
{{EndSpecs}}
* [http://w140.com/tek_067_0587_01.pdf 067-0587-01 Manual (PDF)]
* [http://w140.com/tek_067_0587_01.pdf 067-0587-01 Manual (PDF)]


133

edits