39,660
edits
No edit summary |
(specs, year) |
||
Line 2: | Line 2: | ||
The '''Tektronix P6021''' is a current probe for oscilloscopes. It can be used with the included passive termination, a [[134|type 134 current probe amplifier]], or a [[3A9]] or [[7A14]] current probe amplifier scope plug-in. Bandwidth is up to 52 MHz. | The '''Tektronix P6021''' is a current probe for oscilloscopes. It can be used with the included passive termination, a [[134|type 134 current probe amplifier]], or a [[3A9]] or [[7A14]] current probe amplifier scope plug-in. Bandwidth is up to 52 MHz. | ||
{ | ==Specifications== | ||
{| class="wikitable" | |||
! | |||
! P6021 passive | |||
! P6021 + [[134]] | |||
! P6022 passive | |||
! P6022 + [[134]] | |||
|- | |||
! Sensitivity | |||
| 2 mA/mV or 10 mA/mV || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div) | |||
| 1 mA/mV or 10 mA/mV || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div) | |||
|- | |||
! Bandwidth (-3 dB) | |||
| 60 MHz @ 10 mA/mV || > 40 MHz | |||
| 200 MHz @ 10 mA/mV || > 70 MHz | |||
|- | |||
! LF limit (-3 dB) | |||
| 450 Hz @ 2 mA/mV<br/>120 Hz @ 10 mA/mV || < 12 Hz | |||
| 8500 Hz @ 1 mA/mV<br/>935 Hz @ 10 mA/mV || < 100 Hz | |||
|- | |||
! Maximum current | |||
| colspan="2" align="center" | 15 A<sub>p-p</sub> | |||
| colspan="2" align="center" | 6 A<sub>p-p</sub> | |||
|} | |||
==Manuals== | ==Manuals== |