The Tektronix 7L18 is a spectrum analyzer plug-in introduced in 1978 for 7000-series scopes. It occupies three plug-in bays in the oscilloscope. It covers the range from 1.5 GHz to 18 GHz with the internal mixer and 12.5 GHz to 60 GHz using external mixers. Note that it will NOT do 1.5 to 18 in one span, it will only max span the range selected by the BAND/HARMONIC control. This range is determined by the hetrodyne harmonic number and sign (sideband) and the range of the first LO, which is 2 GHz to 4 GHz.

The input attenuator range is 0 dB to 60 dB in 10 dB steps and is a standard Wenschel Engineering product driven by a belt. The range is encoded for the digital dBm reference level display. The signal then passes through a YIG preselector filter with a 50 MHz bandwidth. After the YIG filter, the signal enters the RF input of the mixer assembly.

The LO is a YIG tuned oscillator covering the 2 GHz to 4 GHz range. It's output passes through a sampler assembly used to PLL the YIG on small resolution bandwidths. The output of the sampler goes to the LO input on the mixer. The mixer IF output is 510 MHz.

Manual needed (with schematics).

Service views:


Boards within the digital card cage (from back to front):

The Micro Computer board has three very old Intel chips: an Intel 4004 processor, an Intel 4002 320 bit RAM and 4 bit output port, and an Intel D4289 standard memory interface. The D4289 connects the 4004 to the two 2Kx8 ROMs. Note that in an earlier version of the 7L18 there were 6 ROMs.

Internal views of the analog parts: