485/Repairs: Difference between revisions
Jump to navigation
Jump to search
(Created page with "==CRT== A healthy Tek 485 CRT has been observed to have a filament current of 98 mA. A broken Tek 485 CRT that lost its vacuum was observed to have a filament current of 153 mA.") |
No edit summary |
||
Line 2: | Line 2: | ||
A healthy Tek 485 CRT has been observed to have a filament current of 98 mA. | A healthy Tek 485 CRT has been observed to have a filament current of 98 mA. | ||
A broken Tek 485 CRT that lost its vacuum was observed to have a filament current of 153 mA. | A broken Tek 485 CRT that lost its vacuum was observed to have a filament current of 153 mA. | ||
[[Category:Instrument repair reports]] |
Latest revision as of 06:42, 1 November 2022
CRT
A healthy Tek 485 CRT has been observed to have a filament current of 98 mA. A broken Tek 485 CRT that lost its vacuum was observed to have a filament current of 153 mA.