A6901: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(NEW in Tek catalog 1981 p.287)
No edit summary
 
(2 intermediate revisions by the same user not shown)
Line 1: Line 1:
[[File:A6901.jpg | thumb | 350px | right]]
{{Instrument Sidebar
The '''Tektronix A6901 Ground Isolation Monitor''' connects between mains power and test equipment  
|manufacturer=Tektronix
without connecting protective ground, in order to safely allow referencing measurements to a low voltage  
|model=A6901
instead of ground, or break ground loops.
|class=Power supply
|series=
|summary=Ground Isolation Monitor
|image=A6901.jpg  
|caption=Tektronix A6901 Ground Isolation Monitor
|introduced=1981
|discontinued=(?)
|designers=
|manuals=
* [[Media:070-3618-00.pdf|Tektronix A6901 Manual]] (PDF)
}}
The '''Tektronix A6901 Ground Isolation Monitor''' connects between mains power and test equipment without connecting protective ground, in order to safely allow referencing measurements to a low voltage instead of ground, or break ground loops.


It monitors the floating equipment ground voltage. If the voltage exceeds 40 V peak,  
It monitors the floating equipment ground voltage. If the voltage exceeds 40 V peak, or ground current exceeds a selectable limit of 0.5, 3.5 or 5 mA, the A6901 disconnects mains power and grounds the test equipment.
or ground current exceeds a selectable limit of 0.5, 3.5 or 5 mA, the A6901 disconnects  
mains power and grounds the test equipment.


{{BeginSpecs}}
{{BeginSpecs}}
Line 16: Line 25:
{{EndSpecs}}
{{EndSpecs}}


==Manuals==
==Links==
{{Documents|Link=A6901}}


* [http://w140.com/tek_a6901.pdf Tektronix A6901 Manual (PDF)]


[[File:A6901 principle.jpg | thumb | 350px]]
[[File:A6901 principle.jpg | thumb | 350px| Operating principle (click to enlarge)]]


[[Category:Power supplies]]
[[Category:Power supplies]]
[[Category:Introduced in 1981]]

Latest revision as of 03:33, 10 October 2023

Tektronix A6901
Ground Isolation Monitor
Tektronix A6901 Ground Isolation Monitor

Produced from 1981 to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix A6901 Ground Isolation Monitor connects between mains power and test equipment without connecting protective ground, in order to safely allow referencing measurements to a low voltage instead of ground, or break ground loops.

It monitors the floating equipment ground voltage. If the voltage exceeds 40 V peak, or ground current exceeds a selectable limit of 0.5, 3.5 or 5 mA, the A6901 disconnects mains power and grounds the test equipment.

Key Specifications

DC trip voltage 28 VRMS (40 V peak) or ±40 V DC
Current trip ±0.5 mA, ±3.5 mA and ±5 mA max. selectable
Trip delay <200 ms
Line voltage 90-128 V; 180-250 V; 48-66 Hz
Load power 500 W max.

Links

Documents Referencing A6901

Document Class Title Authors Year Links
60AX-4741.pdf Application Note Floating Oscilloscope Measurements ... And Operator Protection 1981


Operating principle (click to enlarge)