P6015: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 36: Line 36:


[[Category:1 MOhm Oscilloscope Probes]]
[[Category:1 MOhm Oscilloscope Probes]]
[[Category:High Voltage Probes]]
[[Category:High Voltage Oscilloscope Probes]]

Revision as of 08:40, 6 December 2015

Tektronix P6015 in box

The Tektronix P6015 is a 1000 × high-voltage probe usable up to 20 kV (40 kV peak pulse). It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak). The original filling material, Freon-114 aka Fluorocarbon 114 (R-114, CFC-114), is no longer available due to its ozone-depleting properties. In later models (P6015A), a silicone based replacement is used.

The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants.

Derating curves

Key Specifications

Attenuation ratio 1000 : 1 (variable by 9%)
Input impedance 100 MΩ // 3 pF
Rise time 4.67 ns
Bandwidth 75 MHz
Max. input voltage
  • With dielectric filling: 20 kV DC or RMS to 100 kHz, pulse up to 40 kV (10%, < 0.1s)
  • Without dielectric filling: 13 kV DC or RMS to 500 kHz, pulse up to 18 kV (10%, < 0.1s)
Termination box

Manuals

Links

Pictures