The '''Tektronix Type Z''' is a comparator plug-in for [[500-series scopes]] | The '''Tektronix Type Z''' is a comparator plug-in for [[500-series scopes]], [[introduced in 1960]] and dropped after 1968. | ||
The Type Z has two inputs, A and B, plus a precision comparison voltage, V<sub>C</sub>, that is | The Type Z has two inputs, A and B, plus a precision comparison voltage, V<sub>C</sub>, that is | ||
adjustable from 0 to | adjustable from 0 to ±1 V, 10 V, or 100 V via a ten-turn pot. | ||
V<sub>A</sub>−V<sub>C</sub>, or V<sub>C</sub>−V<sub>B</sub>. | The following differential amplifier can be configured to display V<sub>A</sub>, −V<sub>B</sub>, | ||
V<sub>A</sub>−V<sub>B</sub>, V<sub>A</sub>−V<sub>C</sub>, or V<sub>C</sub>−V<sub>B</sub>. | |||
This allows DC-coupled measurements of small signals with large DC offsets. | |||
Unlike later comparators, | Unlike later comparators, Type Z has separate step attenuators for each input, | ||
permitting the user to view the difference between signals of widely differing amplitudes. | permitting the user to view the difference between signals of widely differing amplitudes. | ||
With a pair of "Disconnect Signal" buttons, the Type Z was the first | It is also the only comparator that can handle ±100 V common mode at full sensitivity. | ||
With a pair of "Disconnect Signal" buttons, the Type Z was the first | |||
movement toward a "GND" position on the input switches. | movement toward a "GND" position on the input switches. | ||
The Type [[P6023]] probe with R and C compensation, [[introduced in 1962]], | The Type [[P6023]] probe with R and C compensation, [[introduced in 1962]], | ||
Line 43: | Line 33: | ||
==Specifications== | ==Specifications== | ||
[[Category:Specifications needed]] | Maximum sensitivity is 50 mV/div, and common-mode rejection at DC and low frequencies is 40,000 to 1. | ||
Risetime as a general-purpose amplifier is 27 ns (13 MHz bandwidth), but it can track a common-mode risetime of 7 ns. | |||
[[Category:Specifications needed]] | |||
==Internals== | |||
The Type Z uses both tubes and transistors, capitalizing on the strengths of each. | |||
The step attenuators are [[turret attenuators]]. | |||
==Links== | ==Links== | ||
Line 51: | Line 50: | ||
<gallery> | <gallery> | ||
Tek z front.jpg| | Tek z front.jpg|3/4 View | ||
Tek z 1.jpg| Front View | Tek z 1.jpg| Front View | ||
Tek z 7.jpg | Front View | Tek z 7.jpg | Front View | ||
Line 70: | Line 69: | ||
Tek_z_schem.png|Main Amplifier | Tek_z_schem.png|Main Amplifier | ||
</gallery> | </gallery> | ||
[[Category:500 series plugins]] | [[Category:500 series plugins]] | ||
[[Category:Differential amplifiers]] | [[Category:Differential amplifiers]] |
Revision as of 18:40, 22 February 2021
Template:Plugin Sidebar 2 The Tektronix Type Z is a comparator plug-in for 500-series scopes, introduced in 1960 and dropped after 1968.
The Type Z has two inputs, A and B, plus a precision comparison voltage, VC, that is adjustable from 0 to ±1 V, 10 V, or 100 V via a ten-turn pot.
The following differential amplifier can be configured to display VA, −VB, VA−VB, VA−VC, or VC−VB.
This allows DC-coupled measurements of small signals with large DC offsets.
Unlike later comparators, Type Z has separate step attenuators for each input, permitting the user to view the difference between signals of widely differing amplitudes.
It is also the only comparator that can handle ±100 V common mode at full sensitivity.
With a pair of "Disconnect Signal" buttons, the Type Z was the first movement toward a "GND" position on the input switches.
The Type P6023 probe with R and C compensation, introduced in 1962, was specifically designed for the Type Z but also marketed for the Type G.
Specifications
Maximum sensitivity is 50 mV/div, and common-mode rejection at DC and low frequencies is 40,000 to 1.
Risetime as a general-purpose amplifier is 27 ns (13 MHz bandwidth), but it can track a common-mode risetime of 7 ns.
Internals
The Type Z uses both tubes and transistors, capitalizing on the strengths of each.
The step attenuators are turret attenuators.