DM43

From TekWiki
Jump to navigation Jump to search
Tektronix DM43
DMM top for 464/465/466/475
DM43 on top of a 463

Compatible with 464

Produced from 1974 to (?)

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix DM43 is a digital multimeter module that attaches to the top of the 463, 464, 465, 466, and 475. The DM43 was later replaced by the DM44. The DM43 is identical to the DM40 except for the added temperature measurement capabilities.

In addition to its voltage, resistance and temperature measurement functions, it can also provide a digital time readout, i.e. the B timebase delay relative to a zero point selected by the user.

Key Specifications

DC V Range 200 mV to 1.2 kV in 5 decade steps
DC V Accuracy 0.1% of reading, ±1 count
DC V Input Resistance 10 MΩ; Removable internal wire strap to increase input resistance to 1000 MΩ on 200 mV and 2 V ranges
Max. Input ±500 V (DC + peak AC) to chassis in 200 mV and 2 V ranges, ±1200 V (DC + peak AC) between positive input and common input in other ranges
Resistance Range 200 Ω to 20 MΩ in 6 decade steps
Resistance Accuracy 200 Ω, 2 kΩ and 20 MΩ, 0.75%, ±1 count; 20 kΩ, 200 kΩ and 2 MΩ, 0.30%, ±1 count
Time Delay Accuracy within 1.5% ±1count for 475, within 2.5% ±1 count for 464, 465, 466
Temperature Range −55 °C to +150 °C (with P6430 probe)
Temperature Accuracy −55 °C to +125 °C: within 2 °C

Links

Documents referencing DM43

Document Class Title Authors Year Links
Tekscope 1974 V6 N4.pdf Article The Oscilloscope with the Digital Multimeter Dennis Bratz 1974
Tekscope 1975 V7 N1.pdf Article Sweep Delay Today Dave Dunlap 1975
Tekscope 1976 V8 N1.pdf Article Delta Time Measurement for the 7000 Series Paul Farley Les Larson Bruce Hofer 1976

Pictures

Components

Some Parts Used in the DM43

Part Part Number(s) Class Description Used in
Siliconix LD110/111A 156-0476-00 156-0477-00 156-1268-00 Monolithic integrated circuit 3½ digit A/D converter DM40 DM43 DM44 DM502 DM502A DM505