1502/Repairs: Difference between revisions

Jump to navigation Jump to search
m
Error in capacitor reference
mNo edit summary
m (Error in capacitor reference)
Line 34: Line 34:
Often, the internal plastic stops get broken from operators forcing the dial past its intended stop, rendering the part useless.
Often, the internal plastic stops get broken from operators forcing the dial past its intended stop, rendering the part useless.


On the original CRT 1502, complete loss of trace may be caused by corrosion/ failure of C6525/C6429 which supply raw DC to the +10V regulator.  
On the original CRT 1502, complete loss of trace may be caused by corrosion/failure of C6341/C6246 (100uF 30V) which supply raw +25V DC.




[[Category:Instrument repair reports]]
[[Category:Instrument repair reports]]
13

edits

Navigation menu