1502/Repairs: Difference between revisions

Jump to navigation Jump to search
m
no edit summary
(Created page with "There are four key parts in the 1502 that make repair difficult. The original sampling gate (CR1732, Tek 152-0631-00) is no longer available, but Avago RF Schottky barr...")
 
mNo edit summary
Line 33: Line 33:
Another common failure is a defective DISTANCE dial – operators expect the numerical readout of this dial to go from zero to 999 and it is not designed to do that.   
Another common failure is a defective DISTANCE dial – operators expect the numerical readout of this dial to go from zero to 999 and it is not designed to do that.   
Often, the internal plastic stops get broken from operators forcing the dial past its intended stop, rendering the part useless.
Often, the internal plastic stops get broken from operators forcing the dial past its intended stop, rendering the part useless.
On the original CRT 1502, complete loss of trace may be caused by corrosion/ failure of C6525/C6429 which supply raw DC to the +10V regulator.




[[Category:Instrument repair reports]]
[[Category:Instrument repair reports]]