13
edits
(Created page with "There are four key parts in the 1502 that make repair difficult. The original sampling gate (CR1732, Tek 152-0631-00) is no longer available, but Avago RF Schottky barr...") |
mNo edit summary |
||
Line 33: | Line 33: | ||
Another common failure is a defective DISTANCE dial – operators expect the numerical readout of this dial to go from zero to 999 and it is not designed to do that. | Another common failure is a defective DISTANCE dial – operators expect the numerical readout of this dial to go from zero to 999 and it is not designed to do that. | ||
Often, the internal plastic stops get broken from operators forcing the dial past its intended stop, rendering the part useless. | Often, the internal plastic stops get broken from operators forcing the dial past its intended stop, rendering the part useless. | ||
On the original CRT 1502, complete loss of trace may be caused by corrosion/ failure of C6525/C6429 which supply raw DC to the +10V regulator. | |||
[[Category:Instrument repair reports]] | [[Category:Instrument repair reports]] |