* cleaned with isopropanol
* cleaned with isopropanol
* airfree filled with oil
* airfree filled with oil
Asembling the probe body:
* put everything together (gold cap, resistor)
* hold the probe body upright
* put the sealing ring on top of the probe body
* fill the probe body until the oil spills over the sealing ring
* put the probe togehter
* clean up the mess


The compensation box is modified as follow:
The compensation box is modified as follow:
* C1: 47 pF ceramic capacitor parallel
* C1: 47 pF ceramic/polystyrene capacitor parallel
* C2: 220 pF ceramic capacitor parallel
* C2: 220 pF ceramic/polystyrene capacitor parallel
* C3: 100 pF polystyrene capacitor parallel
* C3: 100 pF polystyrene capacitor parallel
Compensation:
* follow the rules in the manual
* the probe cable should be stretched out, because it has an importend influence on the first 100 ns


As a rule:  
As a rule:  