DM501A/Repairs: Difference between revisions

Jump to navigation Jump to search
m
501A from DM501/Repairs
No edit summary
m (501A from DM501/Repairs)
 
Line 12: Line 12:


As the PT-100 is connected with only 2 wires, each sensor has to be trimmed with R1001 (offset).
As the PT-100 is connected with only 2 wires, each sensor has to be trimmed with R1001 (offset).
==Switch Repair==
The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches.
Refer to the the pictures and the [[260-1971-00/Repairs| repair]] section of switch assembly [[260-1971-00]].
==R10 ([[307-1015-00]] input voltage divider) Repair==
Fault pattern:  V<sub>DC</sub>, V<sub>AC</sub> and resistance (exept the lowest range) are 2% to low.
Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance.
Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath.
==See Also==
* [[Pushbutton switches]]
==Pictures==
<gallery>
Tek-DM501A-rep-switch.jpg| Replace the push button plunger and spring contacts on a DM501A
Tek-1260-1971-00-rep4.jpg| Defect push button plunger on top and new one below
</gallery>


[[Category:Instrument repair reports]]
[[Category:Instrument repair reports]]
878

edits

Navigation menu