==Supplies out of regulation==
:'''Symptom:''' Several supplies out of regulation
:'''Root cause:''' dry HV electrolytics
:'''Solution:''' Rebuild caps using modern replacements
:<gallery>
Electrolytic repair 1.jpg | Remove defective cap
Electrolytic repair 2.jpg | Cut can near base (using a small hacksaw or a pipe cutter), remove guts
Electrolytic repair 3.jpg | Fit modern cap
Electrolytic repair 4.jpg | Epoxy cap onto base
</gallery>
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:45, 4 May 2019 (PDT)
==HV low and trace invisible==
:'''Symptom:''' No trace, HV low despite maximum setting
:'''Root cause:''' HV [[Bumble Bee capacitors|PIO "Bumble Bee" capacitors]] leaky (~180 μA at 2 kV); no trace because CRT grid too negative wrt. cathode
:'''Solution:''' Replace caps with modern ceramic equivalents
:<gallery>
585 HV caps old.jpg | old PIO caps in place
Black Beauty caps.jpg | old PIO caps removed
585 HV caps new.jpg | new ceramic caps
</gallery>
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:45, 4 May 2019 (PDT)
==No trace==
:'''Symptom:''' Trace invisible
:'''Root cause:''' Wiper of focus pot not in contact
:'''Solution:''' Disassemble focus pot, clean and re-seat wiper
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:45, 4 May 2019 (PDT)
==Trace disappearing==
:'''Symptom:''' Trace fading after ~10 minutes operation
:'''Root cause:''' HV resonant cap (C808) leaky, overheating (visible bulge)
:'''Solution:''' Replace cap with modern ceramic equivalent
:<gallery>
585-c808-therm.jpg | Thermal image
585-c808.jpg      | bulging cap
</gallery>
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:45, 4 May 2019 (PDT)
==Time base inoperative==
==Time base inoperative==
:'''Symptom:''' No trace
:'''Symptom:''' No trace
Line 4: Line 41:
:'''Solution:''' Solder replacement lead after grinding away some glass (or replace neon)
:'''Solution:''' Solder replacement lead after grinding away some glass (or replace neon)
:<gallery>
:<gallery>
585-neon-1.jpg
585-neon-1.jpg | broken lead
585-neon-2.jpg
585-neon-2.jpg | repaired leads (before cleaning and epoxy encapsulation)
</gallery>
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:45, 4 May 2019 (PDT)
 
==Wavy trace ==
:'''Symptom:''' Wavy trace
:'''Root cause:''' C786 (100 µF 25 V) has high internal resistance (ESR ~ 20 Ω) causing +12.6 V regulator circuit to oscillate
:'''Solution:''' Replace C786 with new part.
 
:<gallery>
585 wavy trace.JPG | wavy trace
585_wavy_trace_new_electrolytics.JPG |  +12.6 v PSU with  new eletrolytics.
</gallery>
</gallery>
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:16, 4 May 2019 (PDT)
:--[[User:Matt|Matt]] ([[User talk:Matt|Matt]]) 08:53, 5 May 2019 (PDT)
 
 
[[Category:Instrument repair reports]]