The '''Tektronix 5000-series (5K)''' was introduced in 1971 as a series of lower-cost laboratory oscilloscopes with versatile plug-ins. They generally have larger display screens than the more expensive and higher performance [[7000-series scopes|7000-series]], but have very low bandwidth in comparison − 51xx mainframes are limited to 2 MHz, 54xx series offer 60 MHz. | |||
== The Basics of the 5000-Series == | == The Basics of the 5000-Series == | ||
5000-series mainframes consist of an acquisition unit (the [[5103]] or [[5403]]) containing | |||
the plug-in compartments and low-voltage power supply, and a display unit (e.g., the | |||
[[D10]], [[5111 | D11]], [[5112 | D12]], [[D13]], etc). This design differs from most other Tek equipment of the era except the [[7704A]] from the 7000 series. | |||
The idea was that the display unit could be sold separately for custom integrated uses by other manufacturers at a lower cost than the [[:Category:600_series_CRT_monitors|600-series monitors]], and at the same time provide | |||
a larger display area. | |||
Within the Tektronix product line, the D10 and D11 display units (non-storage and bi-stable storage, respectively) are also used in constructing the [[577|577 semiconductor curve tracer]], | |||
===Nomenclature === | |||
Within the 5000-series, however, this design led to documentation problems as each oscilloscope | |||
ended up with two model numbers and two serial numbers. | |||
In 1977, Tek began to eliminate the serial number from the display unit for the scopes that | |||
were listed as “5103/D13” for example, using the serial number from the display unit for the complete oscilloscope. | |||
Along with this change, Tek began referring to the various models with the acquisition unit | |||
prefix followed by the “D” number, such as [[5110]] for a 5103/D10, [[5113]] for a 5103/D13 | prefix followed by the “D” number, such as [[5110]] for a 5103/D10, [[5113]] for a 5103/D13 | ||
or [[5441]] for a 5403/D41. Thus, documentation became much simpler. | or [[5441]] for a 5403/D41. Thus, documentation became much simpler. | ||
Line 33: | Line 35: | ||
manual, you have all the operation and service information you need for that model. | manual, you have all the operation and service information you need for that model. | ||
It will be pointless and take a lot more time, energy and money to find separate | It will be pointless and take a lot more time, energy and money to find separate | ||
manuals for a | manuals for a 5103 and a D13 than it will be to find a 5113 manual. | ||
== Mainframe | === Mainframe Numbering Convention === | ||
While the 7000-series has a certain convention for numbering the mainframes, the | While the 7000-series has a certain convention for numbering the mainframes, the | ||
5000-series has very little convention. The | 5000-series has very little convention. The '''51-''' and '''54-''' prefixes indicate the respective series. | ||
After that, there is no numbering convention - unlike the 7000s, all 5xxx mainframes have three compartments, which eliminates any special meaning that would be attributed to the last digit of the model number. | |||
convention | |||
special that would be attributed to the last digit of the model number. | |||
The | The -10, -12, -13, -14, -15 and -16 suffixes have no special meaning at all. Some models are non-storage, | ||
some are bi-stable storage, some are dual-beam, some are both dual-beam and storage: | |||
{| class="wikitable sortable" | |||
|- | |||
! Model !! Beams !! Storage !! Notes | |||
|- | |||
| [[5110]] || Single beam || non-storage || | |||
|- | |||
| [[5111]] || Single beam || split-screen bi-stable || | |||
|- | |||
| [[5111A]] || Single beam || split-screen bi-stable || Updated circuitry with 50 Div/µs writing rate | |||
|- | |||
| [[5112]] || Dual beam || non-storage || | |||
|- | |||
| [[5113]] || Dual beam || split-screen bi-stable storage || 20 Div/µs writing rate | |||
|- | |||
| [[5114]] || || || ''Aha! Tektronix never made a 5114!'' <br>Who knows why they skipped over it to the 5115? | |||
|- | |||
| [[5115]] || Single beam || split-screen bi-stable storage || >200 Div/µs writing rate | |||
|- | |||
| [[5116]] || Single beam || || tri-color display using an LCD shutter | |||
|- | |||
| [[5223]] || Single beam || digital storage || | |||
|- | |||
| [[5440]] || Single beam || non-storage || | |||
|- | |||
| [[5441]] || Single beam || variable-persistence storage || | |||
|- | |||
| [[5444]] || Full dual beam || || Independent horizontal deflection systems using the [[5B44]] timebase | |||
|} | |||
(Option 3 enhances the writing rates to 200 Div/µs on most 5100-series storage | (Option 3 enhances the writing rates to 200 Div/µs on most 5100-series storage | ||
mainframes.) | mainframes.) | ||
=== Plug-In Numbering Convention === | |||
5000-series plug-in numbering coincides more with that of the 7000-series. The | 5000-series plug-in numbering coincides more with that of the 7000-series. The | ||
first digit, a “5”, designates the plug-in as being for use in a 5000-series | first digit, a “5”, designates the plug-in as being for use in a 5000-series | ||
Line 78: | Line 93: | ||
== The Two Faces of 5K and Compatibility == | == The Two Faces of 5K and Compatibility == | ||
The 5000-series is divided into two | The 5000-series is divided into two product lines, the low-frequency | ||
5100-series and the high-frequency 5400-series. The 5100-series has a maximum | 5100-series and the high-frequency 5400-series. The 5100-series has a maximum | ||
mainframe bandwidth of 2 MHz and lower bandwidths may be inflicted with plug-ins | mainframe bandwidth of 2 MHz and lower bandwidths may be inflicted with plug-ins | ||
such as the [[5A22N]]. The 5400-series | such as the [[5A22N]]. The 5400-series has a maximum mainframe bandwidth of 60 MHz, | ||
again this figure is at the mercy of lower-bandwidth plug-ins. | again this figure is at the mercy of lower-bandwidth plug-ins. | ||
Line 91: | Line 106: | ||
this series. | this series. | ||
Four plug-ins of the 5000-series directly mirror “twins” in the 7000-series: the | |||
[[5CT1N]] ([[7CT1N]]) curve tracer, [[5A22N]] ([[7A22]]) differential amplifier, | |||
[[5A13N]] ([[7A13]]) differential comparator, and the [[5S14]] ([[7S14]]) sampler. | |||
Front panel layout and function of these “twins” are nearly identical, | |||
although the bandwidth of the 7A13 is significantly higher than its 5K counterpart. | |||
Note also that the 7000 plug-in interface is significantly different from the 5000 series, | |||
so the plug-ins are not identical but may use common parts together with adapters, | |||
like in the 7S14 that [[Media:Tek-7s14-right.jpg|contains the full 5S14 board]]. | |||
===Nomenclature === | |||
“N” at the end of any Tektronix model number indicates a unit that ''does not'' have | “N” at the end of any Tektronix model number indicates a unit that ''does not'' have | ||
on-screen readout capability. All of the low-bandwidth plug-ins carry that designation, | on-screen readout capability. All of the low-bandwidth plug-ins carry that designation, | ||
Line 96: | Line 121: | ||
ever have the “N” designation even though they have no readout capability. | ever have the “N” designation even though they have no readout capability. | ||
All 7000-series products without readout capability have this designation on | All 7000-series products without readout capability have this designation on | ||
everything: [[7403N]], [[7603N]], [[7A18N], [[7B53AN]], etc. It is not known why the 5110, for | everything: [[7403N]], [[7603N]], [[7A18N]], [[7B53AN]], etc. It is not known why the 5110, for | ||
instance, is not the 5110N. One must suppose that it's simply because ''none'' of the | instance, is not the 5110N. One must suppose that it's simply because ''none'' of the | ||
mainframes ever have or can have readout capability. But this doesn't explain why | mainframes ever have or can have readout capability. But this doesn't explain why | ||
Line 103: | Line 128: | ||
option for use in the 5400-series mainframes. | option for use in the 5400-series mainframes. | ||
===Compatibility === | |||
5100-series plug-ins may be used in | 5100-series plug-ins may be used in | ||
''any'' (5100- or 5400-series) mainframe in the 5K series. They will severely limit | ''any'' (5100- or 5400-series) mainframe in the 5K series. They will severely limit | ||
Line 126: | Line 146: | ||
At this point, one may be curious and wonder why the 5400-series plug-ins cannot be | At this point, one may be curious and wonder why the 5400-series plug-ins cannot be | ||
used in the 5100-series mainframes. A close look at the interface connector on | used in the 5100-series mainframes. A close look at the [[5000 Series plug-in interface|interface connector on | ||
both series will show that the 5400-series has two extra power supply voltages, | both series]] will show that the 5400-series has two extra power supply voltages, | ||
±15 V on interface connector pins 6A and 6B. 5400-series plug-ins will not find those | ±15 V on interface connector pins 6A and 6B. 5400-series plug-ins will not find those | ||
two supply voltages in a 5100-series mainframe, so they will not operate correctly. | two supply voltages in a 5100-series mainframe, so they will not operate correctly. | ||
Line 137: | Line 157: | ||
The mechanical aspects of the TM500 line of modular instrumentation and those of | The mechanical aspects of the TM500 line of modular instrumentation and those of | ||
the 5000-series are identical except for the interface connector keying. | the 5000-series are identical except for the interface connector keying, however, | ||
been an instance of a TM500 frequency counter installed in a 5000-series mainframe | '''TM500 instruments and 5000-series instruments are electrically totally incompatible with each other, and any attempt to operate one type of plug-in in the other type of mainframe may result in the destruction of the plug-in, mainframe or both.''' | ||
(There has been an instance of a TM500 frequency counter installed in a 5000-series mainframe | |||
with the LED display bright and presumably active, as the photo illustrated for | with the LED display bright and presumably active, as the photo illustrated for | ||
some items being sold on an on-line auction | some items being sold on an on-line auction, but it appears likely that this image | ||
was | was “photoshopped” by the unscrupulous seller.) | ||
In the 5000-series, all power supply voltages are developed | In the 5000-series, all power supply voltages are developed by the mainframe and | ||
by the mainframe and supplied to the plug-ins as regulated voltages | supplied to the plug-ins as regulated DC voltages of +5 V, ±15 V (5400-series) and ±30 V. | ||
(5400-series) and ±30 V. There are no power supply circuits other than supply | There are no power supply circuits other than supply decoupling and small zener regulators in the plug-ins. | ||
decoupling and small zener regulators in the plug-ins. | |||
On the other hand, the TM500-series power modules (the “mainframes”) supply AC and | On the other hand, the TM500-series power modules (the “mainframes”) supply AC and | ||
Line 158: | Line 173: | ||
are not supplied on similar interface connector pins as the 5000-series. All of | are not supplied on similar interface connector pins as the 5000-series. All of | ||
the power supply regulation control circuitry is inside the individual plug-in for | the power supply regulation control circuitry is inside the individual plug-in for | ||
customized voltages, since each plug-in instrument | customized voltages, since each plug-in instrument is a radical design departure | ||
from the other instruments. The power module also contains one NPN and one PNP | from the other instruments. The power module also contains one NPN and one PNP | ||
power transistor for each compartment for use by the instrument installed as a | power transistor for each compartment for use by the instrument installed as a | ||
Line 190: | Line 205: | ||
a 5A22N and selectively limit the upper frequency cutoff. | a 5A22N and selectively limit the upper frequency cutoff. | ||
Some may think that this low bandwidth makes the 5100-series worthless. In | Some may think that this low bandwidth makes the 5100-series worthless. | ||
reality, there are five things that make the 5100-series attractive: | In reality, there are five things that make the 5100-series attractive: | ||
# Those who work in the low-frequency arena of | # Those who work in the low-frequency arena of audio sound systems, LF and VLF radio, mechanical transducers, hearing aids, etc. will find the lack of HF noise in the 5100-series to be a blessing; | ||
audio sound systems, LF and VLF radio, mechanical transducers, hearing aids, etc. | # The plug-in versatility of the 5100-series can be matched only by that of the 7000-series, and then at a far higher price in cost and complexity; | ||
will find the lack of HF noise in the 5100-series to be a blessing; | |||
# The plug-in versatility of the 5100-series can be matched only by that of the 7000-series, and | |||
then at a far higher price in cost and complexity; | |||
# A large display screen, internal graticule and fine, sharp trace; | # A large display screen, internal graticule and fine, sharp trace; | ||
# Simple and easy troubleshooting and repair, usually involving no custom, complex and hard-to-find | # Simple and easy troubleshooting and repair, usually involving no custom, complex and hard-to-find [[:Category:Tektronix-made integrated circuits|Tektronix-made custom ICs]]; | ||
[[:Category:Tektronix-made integrated circuits|Tektronix-made custom ICs]]; | # Three nice plug-ins, the [[5CT1N]] semiconductor curve tracer, the [[5L4N]] audio spectrum analyzer and the [[5S14N]] sampling system give the 5100-series power that you can't find in any of Tek's portable scopes. The [[5S14N]] increases the effective bandwidth of the system to around 1 GHz. | ||
analyzer and the [[5S14N]] sampling system give the 5100-series power that you can't | |||
find in any of Tek's portable scopes. The [[ | |||
of the system to around 1 GHz. | |||
== Some Interesting Notes == | == Some Interesting Notes == | ||
Line 223: | Line 231: | ||
preamplifier that originally sold for $25 in 1971! Ten years later, the price had | preamplifier that originally sold for $25 in 1971! Ten years later, the price had | ||
risen to $160. | risen to $160. | ||
== History of the 5000 and Subsequent Series of Instruments by [[Ahne Oosterhof]] == | |||
<blockquote> | |||
It was late 1969 when the low frequency scope group was about finished with the [[5030]] scopes (1969), | |||
when [[Val Garuts]] and I prepared the product proposal for what first became to be called the LF-Scopes, | |||
then the [[D10]]/[[5103N]] and finally the 5000-series. | |||
The proposal went two or three times to the Engineering Council before it was accepted (Jan 1970). | |||
The group was formed under leadership of [[Jerry Shannon]]. | |||
I had built a cardboard model of what we envisioned, half the scope for the CRT, etc. | |||
and the other half for three plugins and power-supply, all within 5.25 inch height, | |||
a standard height increment for 19” racks, and the mainframe with three plugins for less than $1000. | |||
Above the plugins a strip of space was envisioned for plugin on/off switching, etc. | |||
I remember [[Larry Weiss]] walking by my bench suggesting that we should make the plug-ins maximum height | |||
and build that switching functionality into the individual plugins. Seemed like a good idea. | |||
One year later (Jan 1971) with about 12 people on the project (EE, ME, manual, ecb) | |||
we had two scopes and three plug-ins in production using several new manufacturing methods | |||
and many new low-cost components. Almost by default I was working | |||
as the lead person on the mainframes and [[Gary Vance]] performed that function for the plug-ins. | |||
Throughout that project some big hurdles had to be overcome, e.g., using tin plated | |||
rather than gold plated circuit boards, soldering active components in the board | |||
rather than using sockets, using low cost open potentiometers rather than fully closed ones. | |||
I had asked the mechanical designers under [[Bud Deibele]] to make the frame weak enough | |||
that it would fail any shake, shock and transportation test. | |||
Then make it a little stronger to barely pass those tests. | |||
The pieces of the half frames were riveted together rather than bolted | |||
and to ship an empty frame required a cardboard insert to prevent distortion of the frame in shipment. | |||
All in the name of lowering costs. | |||
In the assembly area kits for 10 plug-ins were issued to individual assemblers and after finishing these, | |||
the assemblers tested those themselves which led to fewer assembly errors: | |||
smoke or smells signal loudly that you have made a mistake! | |||
Even though it took some effort to convince the manufacturing manager, | |||
[[Al Peecher]], that it would work and he made it a go. | |||
During the design phases Jerry suggested that we make many presentations | |||
throughout the company to Area Representative groups. | |||
That paid off big, because as people learned about our project and our challenges | |||
they were more than willing to provide us with any help we needed. | |||
As we reached the finish of the initial project we looked for product ideas for the future | |||
and that kind of planning always was done as a group effort. | |||
It led to dual beam scopes, additional storage scopes and various plug-ins, | |||
etc. to extend the 5000 series. | |||
We also had noticed that some customers bought display sections only, | |||
to build those into their instrument systems. That became somewhat of a headache, | |||
so we designed a power supply for the displays and turned that idea into a new line of monitors, | |||
an extension of old the 600 series. That replaced the dormant [[601]] and [[602]] displays. | |||
Soon after that we split the group into two business units, scopes and displays. | |||
We also considered a ‘plug-in warmer’ (like in the [[500-series scopes|530/540 series]]) | |||
but that was not really needed as the plugins were completely transistorized | |||
and therefore immediately ready to be used. But it made us consider additional possibilities | |||
like making a multi-meter and other measurement tools and signal source plug-ins. | |||
Soon this turned into a lot more and became the TM500 series (1972?), | |||
using the same design and construction methods as the 5000 series and this line became another separate business unit. | |||
Each of these product lines were successful and stayed in the catalog for about 20 years. | |||
Throughout all this Jerry Shannon encouraged a free flow of ideas; | |||
nothing was out of bounds to be discussed and tried. | |||
As an example, [[Matt Zimmerman]] in his spare times put together a transistor test plugin, | |||
which became the [[5CT1N|5CT1]]. Throughout Jerry liked to take an opposing view of whatever anybody said. | |||
He was willing and able to take any position on any subject from design to politics | |||
and if you thought things were blue he would try to convince you they were red. | |||
I expect he did this to make sure we were strong in our positions and opinions. | |||
In doing so he taught us a lot about dealing with and managing people. | |||
He made it clear to us that we were as design engineers and project leaders | |||
standing in for the owners of the company, as we were making decisions, however small, | |||
that impacted the health of the Corporation. | |||
He also made it clear he expected us to make the right decision more than 80% of the time, | |||
but did not expect 100%, as it would take too long to get there. Risk was allowed. | |||
Yes, he guided us, but in a very unobtrusive way. He let us find our own way, | |||
but was ready to clear a path when needed. In my life I can think of only three people | |||
who I consider my mentors: Professor Oostdijk from my college years, | |||
teaching (forcing?) me to study, Professor [[Dwain Richins]] from the “Tek University”, | |||
teaching me a lot about communicating clearly and in a precise way in the English language and Jerry Shannon. | |||
He taught me a lot about managing people and taking care of unexpected situations. | |||
In my second career, standing alone as entrepreneur/owner/general manager | |||
of a company (A-Laser, Inc.) I have made a lot of use of their teachings. | |||
</blockquote> | |||
A description of most of the 5000 series products with a timeline can be found in this website: | |||
http://i9t.net/5000_scopes.html | |||
[[Category:5000 series mainframes]] |
edits