5T1A: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
(2 intermediate revisions by the same user not shown)
Line 11: Line 11:
|manuals=
|manuals=
* [[Media:070-387.pdf|Tektronix 5T1A Manual]]
* [[Media:070-387.pdf|Tektronix 5T1A Manual]]
* [https://w140.com/tek_fcp/tek_type_5t1a_factory_cal_proc.pdf Tektronix 5T1A Field Recalibration Procedure]
* [[Media:Tek type 5t1a factory cal proc.pdf|Tektronix 5T1A Field Recalibration Procedure]]
* [[Media:050-0419-00.pdf|Modification Note Documenting Replacement of 152-0074-00 Tunnel Diode with 152-0381-00]]
* [[Media:050-0419-00.pdf|Modification Note Documenting Replacement of 152-0074-00 Tunnel Diode with 152-0381-00]]
* [[Media:Tek 5t1a irb no ocr.pdf|Tektronix 5T1A Instrument Reference Book]] (needs OCR)
* [[Media:Tek 5t1a irb no ocr.pdf|Tektronix 5T1A Instrument Reference Book]]
* [[Media:Tek 5t1a cal outline.pdf|Tektronix 5T1A Calibration Outline]] (OCR)
* [[Media:Tek 5t1a cal outline.pdf|Tektronix 5T1A Calibration Outline]] (OCR)
* [[Media:Tek 5t1a preliminary.pdf|Tektronix 5T1A Preliminary Manual]]
* [[Media:Tek 5t1a preliminary.pdf|Tektronix 5T1A Preliminary Manual]]
Line 29: Line 29:


Triggering in the 5T1A is accomplished using five [[1N3129]] 20 mA [[tunnel diodes]].
Triggering in the 5T1A is accomplished using five [[1N3129]] 20 mA [[tunnel diodes]].
==See Also==
* [[040-0390-00]] modification kit improving high frequency trigger performance


==Pictures==
==Pictures==

Latest revision as of 22:35, 9 March 2024

Manuals – Specifications – Links – Pictures

The Tektronix 5T1A is a timing plug-in introduced in 1963 for the 661 sampling oscilloscope. It is an improved version of the 5T1.

The job of the 5T1A is to produce three signals – the sampling pulse, the horizontal sweep, and the blanking signal. The input to the 5T1A is a trigger signal. The trigger can arrive from any of four sources:

  • internally from the trigger pick-off in a 4S1 or 4S2A plug-in
  • externally via a GR-874 connector on the front panel of the 5T1A
  • from the calibration oscillator in the 661, via the multi-pin plug-in connector
  • free-running

Each of these modes is suited to some measurement scenarios. In the case of internal triggering, the signal goes from the trigger pick-off in the 4S1 to a coaxial interconnect that passes through the 661, connecting the sampling unit to the timing unit.

Triggering in the 5T1A is accomplished using five 1N3129 20 mA tunnel diodes.