{{Instrument Sidebar
|manufacturer=Tektronix
|model=067-0508-00
|class=Signal generator
|series=
|summary=Amplitude Calibrator
|image=Tek 067-0508-00 3.jpg
|caption=Tektronix 067-0508-00
|introduced=(?)
|discontinued=(?)
|designers=
|manuals=
* [[Media:Tek 067-0508-00.pdf|Tektronix 067-0508-00 Manual]] (OCR)
* [[Media:Tek 067-0508-00 schematics.pdf|Tektronix 067-0508 Schematics]]
}}
The '''Tektronix 067-0508-00''' Amplitude Calibrator generates a square-wave with two precisely controlled voltages and 50 Ω impedance at both voltages.
 
The calibrator uses a 50 Ω shunt resistance at the output, and a precision DC current source that can be steered toward the output, or toward a dummy current sink transistor. 
The built-in multivibrator controls the steering of the current. 
 
The output is through a [[GR-874 connector]]. The [[067-0515-00|067-0515-00 precision 50 Ω terminator]] is intended to be used with the 067-0508-00.
 