Я4С-89: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(Created page with "{{Plugin Sidebar 2 | title=Я4С-89 | summary=Pulse Generator | image=Я4С-89_front.jpg | caption=Я4С-89 front panel | series=Soviet 7000-series clones| }} The '''Я4...")
 
(Introduced 1978 according to https://eltesta.com/wp-content/uploads/2021/12/Alma_Mater_2021.pdf p.242)
 
(10 intermediate revisions by 2 users not shown)
Line 1: Line 1:
{{Plugin Sidebar 2 |
{{Plugin Sidebar
title=Я4С-89 |
|manufacturer=Vilnius Research
summary=Pulse Generator |
|series=Soviet 7000-series clones
image=Я4С-89_front.jpg |  
|type=Я4С-89  
caption=Я4С-89 front panel |
|summary=Pulse Generator  
series=[[Soviet 7000-series clones]]|
|image=Я4С-89_front.jpg  
|caption=Я4С-89 front panel  
|introduced=1978
|discontinued=(?)
|designers=
|manuals=
* ''please add'' [[Category:Manual needed]]
}}
}}
The '''Я4С-89''' is a plug-in for [[Soviet 7000-series clones]] which contains a pulse generator.
It is intended for use with the sampling plugins in a similar manner to the Tek [[S-52]] pulse
generator head. It can free-run or take a 5 V external trigger signal.


The '''Я4С-89''' is a plug-in which contains a pulse generator. It's intended for use with the sampling plugins
The pulse outputs require an external pulse shaping cable in order to be useful. The cables contain a [[Russian tunnel diodes|1И308К 50 mA tunnel diode]].
in a similar manner to the Tek [[S-52]] pulse generator head. It can free-run or take a 5V external trigger
signal.


From the top, the connectors are:
From the top, the connectors are:
* external 5V (TTL) trigger in
* external 5 V (TTL) trigger in
* sync output, variable level and polarity
* sync output, variable level and polarity
* positive-going pulse output
* positive-going pulse output
Line 28: Line 35:
* delay vernier
* delay vernier
* pulse width
* pulse width
then two screwdriver bias adjustments for the positive-going and negative-going pulses.
then two screwdriver bias adjustments for the positive-going and negative-going pulse outputs.
 
 
The connectors are СР50-73 type, which are almost compatible with BNC connectors but the latch is a tight fit and the centre contact can be unreliable.
 
{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Period | 0.01 - 1ms, or external trigger }}
{{Spec | Period | 10 μs to approx 15 ms, or external trigger }}
{{Spec | Post-trigger delay  | to 200ns}}
{{Spec | Post-trigger delay  | 0 to approx 400 ns }}
{{Spec | Rise time | not yet measured }}
{{Spec | Rise time | Depends on external pulse shaper }}
{{Spec | Pulse width | 0.01-0.2 μs}}
{{Spec | Pulse width | 10 ns to 10 μs in two ranges}}
{{Spec | Amplitude | not confirmed }}
{{Spec | Trigger output amplitude | −5 V to +5 V into 1 MΩ, −1 V to +1 V into 50 Ω }}
{{Spec | Pulse output amplitude | 200 mV or 400 mV depending on pulse shaper }}
{{EndSpecs}}
{{EndSpecs}}


==Links==
==Links==


==Pictures==
==Pictures==
Line 46: Line 55:
Я4С-89_front.jpg | Front  
Я4С-89_front.jpg | Front  
Я4С-89_threequarter.jpg | Threequarter view
Я4С-89_threequarter.jpg | Threequarter view
Я4С-89_pulse_shaper_cables.jpg | Pulse shaper cables schematic
Я4С-89_pulse_outputs.jpg | Top trace trigger output, bottom trace + pulse output, pulse length set to 10ns. Note initial positive bump to trigger tunnel diode, then big negative spike 10ns later to switch it off.
</gallery>
</gallery>
[[Category:Soviet 7000-series clones]]
[[Category:Vilnius Research Institute products]]

Latest revision as of 09:00, 21 February 2022

Vilnius Research Я4С-89
Pulse Generator
Я4С-89 front panel

Compatible with Soviet 7000-series clones

Produced from 1978 to (?)

Manuals
  • please add
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Я4С-89 is a plug-in for Soviet 7000-series clones which contains a pulse generator. It is intended for use with the sampling plugins in a similar manner to the Tek S-52 pulse generator head. It can free-run or take a 5 V external trigger signal.

The pulse outputs require an external pulse shaping cable in order to be useful. The cables contain a 1И308К 50 mA tunnel diode.

From the top, the connectors are:

  • external 5 V (TTL) trigger in
  • sync output, variable level and polarity
  • positive-going pulse output
  • negative-going pulse output

The buttons are:

  • manual trigger
  • external trigger mode
  • period 1/0.1/0.01 ms
  • delay 25/50/100/200 ns (can be added by pressing multiple buttons)
  • pulse width 0.01/0.2 μs

The rotary controls are:

  • external trigger level
  • period vernier
  • sync output amplitude, negative-zero-positive
  • delay vernier
  • pulse width

then two screwdriver bias adjustments for the positive-going and negative-going pulse outputs.

The connectors are СР50-73 type, which are almost compatible with BNC connectors but the latch is a tight fit and the centre contact can be unreliable.

Key Specifications

Period 10 μs to approx 15 ms, or external trigger
Post-trigger delay 0 to approx 400 ns
Rise time Depends on external pulse shaper
Pulse width 10 ns to 10 μs in two ranges
Trigger output amplitude −5 V to +5 V into 1 MΩ, −1 V to +1 V into 50 Ω
Pulse output amplitude 200 mV or 400 mV depending on pulse shaper

Links

Pictures