015-0073-00: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
mNo edit summary
 
(5 intermediate revisions by 3 users not shown)
Line 1: Line 1:
The '''Tektronix 015-0073-00''' is an accessory power supply providing +12.5 V and −12.5 V at 150 mA each.
{{Instrument Sidebar
It is intended for probes, e.g. the [[P6045]] or [[282]].
|manufacturer=Tektronix
|model=015-0073-00
|class=Power supply
|series=
|summary=Accessory Power Supply
|image=Tek 015-0073-00 1.jpg
|caption=Tektronix 015-0073-00
|introduced=1967
|discontinued=(?)
|designers=
|manuals=
* [[Media:070-0636-00.pdf|Accessory Power Supply Manual]]
* [[Media:Tek 015-0073-00 fcp april 1967.pdf|015-0073-00 FCP]]
}}
The '''Tektronix 015-0073-00''' is an accessory power supply providing +12.5 V and −12.5 V at 150 mA each. It is intended for probes, e.g. the [[P6045]]. Later the specification of the supply was changed from ±12.5 V ±5% to ±12.0 V ±1%.


The 015-0073-00 uses a mains frequency (non-SMPS) transformer, with separate secondaries for the +12.5 V supply and the −12.5 V supply.
The 015-0073-00 uses a mains frequency (non-SMPS) transformer, with separate secondaries for the +12.5 V supply and the −12.5 V supply.
Line 7: Line 21:
The regulators use [[MPS6521]] NPN bipolar transistors as error amplifiers.
The regulators use [[MPS6521]] NPN bipolar transistors as error amplifiers.
The regulator output is a Darlington pair formed by a [[MPS6515]] and a [[RCA40250]].
The regulator output is a Darlington pair formed by a [[MPS6515]] and a [[RCA40250]].
==Manuals ==
* [[Media:015-0073-00.pdf | Accessory Power Supply Manual (PDF)]]
* http://www.davmar.org/pdf/Tek015-0073-00.pdf


==Pictures==
==Pictures==

Latest revision as of 11:49, 28 November 2023

Tektronix 015-0073-00
Accessory Power Supply
Tektronix 015-0073-00

Produced from 1967 to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 015-0073-00 is an accessory power supply providing +12.5 V and −12.5 V at 150 mA each. It is intended for probes, e.g. the P6045. Later the specification of the supply was changed from ±12.5 V ±5% to ±12.0 V ±1%.

The 015-0073-00 uses a mains frequency (non-SMPS) transformer, with separate secondaries for the +12.5 V supply and the −12.5 V supply.

Both supplies are regulated independently with zener diodes. The regulators use MPS6521 NPN bipolar transistors as error amplifiers. The regulator output is a Darlington pair formed by a MPS6515 and a RCA40250.

Pictures