The '''Tektronix Type 130''' is a self-contained instrument that measures inductance and capacitance.   
The '''Tektronix Type 130''' is a self-contained instrument that measures inductance and capacitance.   
When measuring inductance, the 130 makes an LC oscillator using the device (inductance) under test
The 130 makes an LC oscillator using the device under test
and a capacitor internal to the 130. When measuring capacitance, the 130 makes an LC oscillator using the device
and a capacitor or inductor internal to the 130. That oscillator's frequency is
(capacitance) under test and an inductor internal to the 130.  The frequency of the resulting LC oscillator is
measured by mixing it down using a fixed reference frequency, generating a pulse at each zero-crossing, and
measured by mixing it down using a fixed reference frequency, generating a pulse at each zero-crossing, and
integrating (low-pass filtering) the pulse train.  This produces a voltage that is proportional to the  
integrating (low-pass filtering) the pulse train.  This produces a voltage that is proportional to the  
Line 24: Line 23:


==Links==
==Links==
* Q+A: Type 130 L-C Meter and S-30 Delta Standards. Service Scope No. 18, Feb 1963
* Q+A: Type 130 L-C Meter and S-30 Delta Standards. [[Media:Service Scope 18 Feb 1963.pdf | Service Scope No. 18, Feb 1963]].